Reasons for inadequate or incomplete imaging techniques

Cosmin Caraiani

1

, Yi Dong

2

, Anthony G. Rudd

3

, Christoph F. Dietrich

4,5

1Department of Medical Imaging, “Iuliu Hatieganu” University of Medicine and Pharmacy, Cluj-Napoca, Romania,

2Department of Ultrasound, Zhongshan Hospital, Fudan University, Shanghai, China, 3Division of Health and Social Care Research, King’s College London, London, United Kingdom, 4Medizinische Klinik 2, Caritas-Krankenhaus Bad Mergentheim, Bad Mergentheim, Germany, 5Ultrasound Department, First Affiliated Hospital of Zhengzhou University, China

Received 27.08.2018 Accepted 02.10.2018 Med Ultrason

2018, Vol. 20, No 4, 498-507

Corresponding author: Prof. Dr. med. Christoph F. Dietrich Medizinische Klinik 2, Caritas Krankenhaus Bad Mergentheim, Uhlandstr. 7,

D-97980 Bad Mergentheim, Germany Phone: 49 (0)7931 - 58 - 2201 / 2200 Fax: 49 (0)7931 - 58 - 2290 E-mail: [email protected]

Introduction

Over the last few decades, advances in imaging have transformed the practice of medicine. The diagnostic process in almost every medical field would be incon- ceivable without imaging. Unfortunately, each imaging technique also has contraindications, limitations, and factors limiting its diagnostic potential. Imaging tech- niques can also be potentially harmful to the patient, as in the case of computed tomography (CT), where ionizing radiation is used.

In this paper radiologists and clinicians focus on fac- tors which make imaging techniques either unrealizable or degrade the image quality to a point in which they become uninterpretable. Progress in imaging techniques has also led to a significant decrease of incomplete or uninterpretable imaging procedures.

Magnetic resonance imaging (MRI)

MRI is a very useful and established imaging tool, which can give information of great diagnostic value. MRI was initially developed for neurological purposes but, as the technique progressed, it became an indispensable diagnos- tic tool in musculoskeletal [1], abdominal [2], and gyneco- logic imaging [3] MRI is increasingly used worldwide. It is estimated that in the last 10 years the number of MRI performed each year has increased by 20% [4]. Over 80 million MR exams are performed each year worldwide [5].

Unfortunately, in daily clinical practice, situations are encountered where MRI is not feasible or unable to pro- vide the required amount of information. Such situations Abstract

Even if imaging has developed considerably during the last decades there still exist several factors which limit its capaci- ties. These factors can either limit the usage of a technique or degrade images making them difficult to interpret. Magnetic resonance imaging (MRI) has, as an absolute contraindication, the presence of metallic devices marked as “MRI unsafe” and metallic foreign bodies close to the eye or vital structures. Claustrophobia and artefacts reduce the application and performance of MRI in a significant proportion of patients. The major disadvantages of computed tomography (CT) are the exposure to ionizing radiation inducing malignancies especially in pediatric patients and the risk of contrast induced allergies and ne- phropathy. Ultrasound is a safe, easily available and low-cost imaging technique without significant side effects for the patient.

Obesity or bloating can severely limit ultrasound capacities.

This paper written by radiologists and clinicians, highlights the main reasons leading to inadequate imaging and points out solutions to avoid inaccurate diagnosis due to incomplete imaging or presence of artifacts.

Keywords: Magnetic Resonance Imaging (MRI); Computed Tomography (CT); Ultrasound (US); contraindications;

artefacts

DOI: 10.11152/mu-1736

(2)

can be classified into: 1) situations when MRI is not feasi- ble due to claustrophobia or to absolute contraindications such as the presence of metallic foreign bodies or devices and 2) situations when MRI offers incomplete informa- tion due to inadequate imaging. The presence of metallic artifacts, uncontrolled breathing or patient movement can lead to image degradation. Lack of patient comfort due to a closed tube or to high acoustic noise may also lead to decrease in image quality. Factors related to the machine or the technique used may limit MRI’s potential of of- fering useful information. Contrast media administration must be avoided in some circumstances, but non-admin- istration of contrast media may diminish the amount of useful information.

1) Situations when MRI is “non-feasible”

Foreign bodies or medical implants which represent an absolute contraindication to MRI Materials used as implants and devices are classified as “MRI safe”, “MRI conditional”, or “MRI unsafe” [6].

Patients with MRI safe devices can be scanned without concern, but it should be noted that this terminology re- fers only to potential side effects that could produce harm to the patient; it does not guarantee that the presence of

“MRI safe” material will not degrade image quality. In practice, a patient with an “MRI safe” device such as a hemostatic clip close to a rectal tumor or fistula can re- sult in an examination that is uninterpretable because of artefacts (fig 1).

The presence of “MRI conditional” devices means that MRI is feasible, under some circumstances, while the presence of “MRI unsafe” materials represents a absolute contraindication. The conditions under which patients with materials considered as “MRI conditional”

can be imaged are required to be stated by the manufac- turer [7]. As a general rule, practitioners should never as- sume the MR compatibility of a device if it is not clearly documented in writing. Even if most modern devices are designed to be MRI-safe, the safety of MRI should not be assumed only on the basis that the implanted device was manufactured recently [8].

A device can be “MRI conditional” on some MRI machines and “MRI unsafe” on other machines. Factors related to the MRI examination such as strength of the magnetic field, maximum spatial gradient magnetic field of the magnet or pulse-sequence parameters can vary sig- nificantly among different MR systems leading to vari- able tolerance of implanted devices. Information on the safety of implants and devices can be obtained on the site www.MRIsafety.com [9].

The presence of MRI unsafe materials is a concern to the health system as many patients with MRI unsafe

materials need to undergo an MRI during their lifetime.

Most often, in daily practice, the following devices are MRI unsafe: pacemakers and implantable cardioverter- defibrillators (ICD), implanted neurostimulators, brain aneurysm clips made from ferromagnetic material, hemodynamic monitoring and temporary pacing devic- es; hemodynamic support devices. The frequency with which these devices are present in those who require an MRI during their lifetime has led to a reconsideration of contraindications, ex: 2 million US citizens and 6 mil- lion worldwide have ‘MRI unsafe’ pacemakers. A recent multicenter study performed on patients with “non-MRI conditional” cardiac pacemakers and ICD’s on 1.5T ma- chines involved 1500 patients and no deaths, ventricu- lar arrhythmias, device or lead failure were noted. Of note in this study all devices were reprogrammed before MRI and no MRI’s of the thoracic region were under- taken [10]. However, MRI of such patients, if absolutely needed, has to be performed in experienced centers under very strict conditions of surveillance and machine set- tings. A large population based study estimated that up to 84% of patients with neurostimulation devices (which are, in their vast majority, marked as MRI unsafe) will have indications for an MRI scan during the following 5 years, data which can be somehow extrapolated also to patients with pacemakers or ICD’s [4].

Metal fragments near the eye or great vessels Shifting of metallic foreign bodies under the influ- ence of MRI can lead to damage to vital structures such as nerves, vessels, or the eye. If there is any doubt about the presence or location of a metallic foreign body, an X-ray should be performed prior to the MRI [11]. Sur- gery to remove the foreign body may be needed to safely perform MRI.

In particular cases, when information obtained by MRI is very important for the patient, MRI can be done despite the presence of presumed ferromagnetic foreign bodies – but only if they are not located in or near a vital structure.

Fig 1. MRI of the foot of a football player with an ancient, treated metatarsal fracture: Even if the osteosyntetic material was marked as “MRI-safe” images obtained at the site of the fracture are uninterpretable.

(3)

Anxiety and claustrophobia

Even if it is not a contraindication for MR imaging, claustrophobia and anxiety related reactions may lead to failure of MR imaging. Anxiety due to MRI was compa- rable, to anxiety before surgery [12]. The rate reported for failed examination because of anxiety varies in the literature between 0.7 and 20% [13,14]. It has been re- ported that 4%–30% of the population suffered from anx- iety symptoms during MR examinations and that 3%–5%

of patients were unable to complete MR examinations [14]. Most of the papers refer only to patients who are un- able to complete their MR procedure due to development of anxiety and claustrophobic symptoms, but did not in- clude patients who completely refuse MRI. One study found that 10.6% of woman refused breast MRI due to fear of claustrophobia symptoms [15]. The percentage of patients unable to complete MR procedures can be re- duced to only 0.7% of patients by using good communi- cation techniques, quieter modern scanners with a shorter bore and reducing scanning times to a minimum. Open scanners can also significantly contribute to a reduction in claustrophobia sensations [16]. Modern scanners are open at both ends which offer the option of scanning the patient head-first. Claustrophobic reactions may also be minimized by sedating the patient before the procedure.

Despite all these options a further 0.96% of patients, will have uninterpretable images due to motion artefacts even if they are able to finish the MR examination [17].

2) Situations when MRI is feasible but difficult to interpret/uninterpretable

Presence of artefacts

(related to metallic objects or motion)

The presence of artefacts can lead to poor quality im- ages and difficult interpretation (fig 2).

Metallic artefacts can lead to non-visualization or bad visualization of a body part (fig 3). Methods for reducing the amplitude of metallic artifacts will vary from simple to complex. Some of the simple methods are: a high read- out bandwidth; swapping frequency and phase; using thin sections; lower magnet strength (artifacts are more significant at 3.0T compared to 1.5T); increase the ma- trix; increase the number of excitations (NEX); shorter echo-spacing; the use of metal correction artifact soft- ware such as MAVRIC or SEMAC [18,19].

The cooperation of the patient is needed if MRI is to be properly performed, with the patient being required to remain still during the examination (the duration of such an MRI examination can be up to 30-40 min). Respirato- ry motion is also a major cause of artefacts in abdominal and thoracic imaging. Some sequences require the patient to hold their breath for almost 20 s, a feat that is practical-

ly impossible in severely ill and sedated patients. Either breath-hold techniques or free-breathing with respiratory gating can be used to avoid respiratory related artefacts [20]. If gated acquisitions are to be efficient and artefacts avoided, techniques involving free-breathing require the patient to breath in a regular manner. Faster sequences can also reduce the amplitude of breathing artefacts. Be- fore improvements in MR techniques, including the use of faster sequences, 5% of abdominal MRI were severely degraded due to breathing artefacts. This figure has now dropped to 0.5-1% [21,22].

There are a series of measures which can be taken to minimize motion artifacts: stabilization of the patient;

using surface coils confined to the area of interest; using Fig 2. Brain MRI of a young patient. Severe artifacts due to the presence of a hairclip. Images of the right cerebral hemisphere are uninterpretable. The examination was continued after re- moval of the hairclip.

Fig 3. Brain MRI of a 55-years old female patient with mild headaches in her medical history, asymptomatic in the moment of the examination. Image 3a shows hyperintensity of the bi- lateral posterior cerebral sulci on the Flair sequence, an aspect which is highly suspicious for subarachnoid hemorrhage. Due to the contradiction between the imaging aspect and the pa- tients good condition she was recalled for an repeat examina- tion which proved to be normal (Image 3b). The aspect in 3a was due to artifacts from a hairclip which was not removed by the radiographer.

(4)

spatial saturation pulses; flow saturation pulses; increas- ing number of signals averaged; swapping frequency and phase encoding directions; use of single-slice ultrafast sequences; using self-correcting sequences like PRO- PELLER and BLADE.

3) Administration of Gadolinium based contrast agents (GdCA)

Allergic reactions to Gadolinium contrast agents The rate of acute allergic reactions after administra- tion of GdCA is low, 0.07%-0.7%, with anaphylaxis due to GdCA occuring in approximately 1:100 000 patients [23,24]. The side effects of GdCA are generally mild and include headache, vomiting, nausea, rash, local burning and thrombophlebitis at the site of the injection [25].

Patients with previous allergic reactions to Gadolinium are at a higher risk of developing adverse reactions to subsequent administration of GdCA. Injection of contrast media in such patients should therefore be avoided or, if there is an absolute need for injecting contrast media, pa- tients should be premedicated with corticosteroids and/or antihistamines [23].

Nephrogenic systemic fibrosis (NSF)

NSF is a rare life-threatening disease that affects patients with renal failure who have had GdCA admin- istered; that is why patients with a glomerular filtration rate (GFR) >30 ml/min should not receive these contrast agents. Contrast agents may be classified as high risk (ga- dopentetic acid), medium risk (gadobenic acid) or low risk (gadoteridol) [7], 78% of NSF being associated with the administration of Omniscan (gadodiamide) and 20%

with Magnevist (gadopentetate dimeglumine). Virtually no cases of NSF have been reported after the administra- tion of MultiHance or Eovist [26,27]. Recent data sug- gest that class II Gd-based contrast agents can be safely administrated in patients with impaired renal function.

Precautions still need to be taken and the lowest possible quantity of contrast necessary for the diagnosis should be injected.

Computed tomography (CT)

CT has, over MRI, the advantages of being more eas- ily available and quick to perform. The ability to scan multiple regions and organs in a single examination and in very short period of time (1-3 min) makes CT the tech- nique of choice for the follow-up of oncological patients.

CT is also indicated in abdominal, thoracic, neurological, skeletal pathology and in traumatic and non-traumatic emergencies [28].

There are no absolute contraindications for unen- hanced CT examination, although there is still a patient

risk due to radiation exposure. According to the ALARA (“as low as reasonably achievable”) principles, exposure to the patients must be minimal. Every time a procedure using ionizing radiation can be substituted with a non- ionizing procedure allowing the same amount of infor- mation to be obtained and having the same diagnostic benefit, the non-irradiant procedure should be used. In order to avoid unnecessary irradiation, the radiologist must discuss, on a case-by-case basis, the appropriate- ness of CT scanning.

CT is in some cases inadequate. This may be due to the presence of artefacts. Non-administration of contrast media will decrease the amount of information which can be obtained and there are strong relative contraindica- tions to contrast media administration.

1) Avoiding unnecessary irradiation of patients

Abdomino-pelvic CT is a relatively high-radiation dose examination in comparison with chest and head CT- due to the large number of radiosensitive organs in the field of view (imparts a dose of 6–10 mSv to the adult patient [29]). The mean dose in abdominal CT (8 mSv) corresponds to approximately 11 conventional abdomi- nal radiographs (0.7 mSv), which is equivalent to 3-4 years of natural background radiation [30]. In the United States, radiation exposure from radiological procedures grew from 0.54 mSv per person in 1980 to 3.0 mSv per person in 2006, particularly due to the development of CT. In 2006, CT accounted for one half of all radiation exposure, despite accounting for only 17% of all imaging procedures [31].

Many low-radiation dose protocols have been devel- oped, although most of them also lead to reduced image quality and conspicuity [32,33]. Such protocols are par- ticularly useful in patients <40 years or in conditions in which fine detail is not needed, such as in nephrolithi- asis. Also, the number of scans before and after contrast media injection has to be limited to a minimum neces- sary. Guidelines, such as the American College of Ra- diology (ACR) Appropriateness Criteria, should be used to choose the most appropriate imaging procedure and reduce unnecessary irradiation.

A large population based survey showed that 56% of the medical tests which are overordered are imaging pro- cedures, mostly CT-scans [34]. The use of a Web-based computerized radiology entry system based on the ACR appropriateness criteria have reduced by 3% the outpa- tient CT procedure volume in a single institution [35].

Different papers suggests that CT is especially overused in clinical scenarios such as the diagnosis of minor head trauma [35] (37% of patients have no indication) and the diagnosis of appendicitis [36].

(5)

2) Situations in which CT is indicated but it provides incomplete information

Artefacts in CT and uninterpretable CT images Artefacts can seriously degrade the quality of CT im- ages, in some cases to the point of making them unin- terpretable. Fortunately, the incidence where image al- teration is so profound that interpretation is impossible is low and useful information can be obtained, even from an imperfect examination.

Artefacts can be classified into physics-based arte- facts such as beam-hardening, which are related to the process of image acquisition, patient-related artefacts due to patient movement or metallic objects inside or on the skin of the patient and scanner-based artefacts related to imperfections in scanner functioning [37]. Most of the physics and scanner-based artefacts can be avoided or minimized by the correct use of the technique or by using the software available on new scanners to improve image quality and conspicuity.

Metal artefacts are extremely common, having been documented as affecting up to 10% of scans in one series, although most of them have no major impact on image visualization [38]. However, in some cases the presence of metallic objects can lead to severe streaking artefacts.

To avoid these, patients should be asked to remove pieces of jewelry before the procedure. Furthermore, medical objects such as wires or leads should be removed from the body before the procedure. Gantry angulations can sometimes be used to exclude non-removable objects from the scanned field. If it is not possible to exclude metallic objects from the scanned field, it may still be possible to obtain interpretable images by increasing the kilovoltage and obtaining thin sections; corrections which unfortunately also lead to an increase in radiation dose. In addition, the impact of artefacts on image quality can be reduced by applying software corrections during the post-processing of the images [37]. In recent years, new post-processing techniques have been used in order to reduce metallic artefacts (MAR- metal artefact reduc- tion techniques). They have different commercial names depending on the different vendors- OMAR (Philips), SEMAR (Toshiba) or MARS (General Electric) [39].

The reported reduction of noise by using these softwares is between 53 and 77%.

Motion can cause blurring and double images, as well as long streaks. Faster scanners have considerably reduced motion artefacts [40].

Cardiac CT is particularly prone to motion artifacts and as a result frequently leads to uninterpretable images.

With the increase in speed of the scanners there has been a decrease in the number of uninterpretable exams. Origi-

nally with 4-slice scanners the percentage of uninterpret- able studies was 20 and 40% but with the use of 64 and 128 slice scanners less than 3% of scans are uninterpret- able [41,42].

Contraindications to contrast media administration Contrast media administration is particularly useful in abdominal pathology. In contrast to chest CT, where the presence of air and bone provides a natural contrast, most abdominal structures have a similar soft tissue den- sity, so differentiating between them can be difficult on non-enhanced scans. Additionally, the visualization and correct characterization of focal liver lesions requires ad- ministration of contrast media [43].

Allergy to iodinated contrast media is rare,0.2 to 0.6% of patients developing allergic reactions and 0.04%

developing severe allergic reactions [44,45]. Nearly all life-threatening reactions to contrast media occur in the first 20 minutes after injection.

Patients with prior allergy to contrast media have 5-times greater risk of developing allergic reactions when contrast media is injected a second time. One recent study estimated that 19.5% of patients with an initial moderate to severe allergic reaction to contrast media will develop allergic reactions when contrast media is reinjected [46].

Patients with unrelated allergies have 2-3-fold higher risk of developing allergic reactions to contrast media.

The use of non-ionic contrast media has contributed to a significant decrease in adverse reactions during contrast media administration [47]. Infants and the elderly have lower risk of developing allergic reactions to contrast media administration comparing with middle-aged peo- ple. The pretesting of patients to contrast media does not predict which patients are prone to allergic reactions, and does not lower the incidence of contrast media reactions.

In a meta-analysis of over 300 000 cases, the predictive value and sensitivity of pretesting for anticipating the re- action to contrast media were extremely low, with values of 1.2% and 3.7% respectively [48]. Studies have shown that a test-reaction to contrast media can actually increase the incidence of contrast media reactions [49]. Mortality due to low-osmolarity contrast media is extremely low.

A large Japanese study described no fatal reactions in 170 000 injections [47].

History of a prior allergy to contrast media is consid- ered a strong relative contraindication to contrast media administration. If contrast media is absolutely necessary for diagnostic purposes, it can be performed after pre- medication with corticosteroids and antihistamine drugs.

Changing the contrast medium has also lowered the inci- dence of repeated allergic reactions in patients with prior allergic reactions to contrast media administration [50].

Routine premedication for patients with other types of

(6)

allergies is not indicated. Contrast media administration should not be withheld in patients with allergies, other than those to iodinated contrast media.

Even if renal insufficiency is a strong relative con- traindication for contrast media administration anuric patients in end-stage renal disease with no transplanted kidneys can have contrast media safely.

3) Challenges of CT and MRI of the obese patient Severe obesity can also be a limitation for CT and MRI. Most CT and MRI machines cannot accommodate patients with a weight above 160 kg or an abdominal di- ameter of more than 60 cm [51], although some modern machines can exceed these limits up to 190-200 kg. For both techniques, if patients fit into the gantry the images will generally be diagnostic. Large-weight capacity CT and MRI machines capable of imaging patients heavier than 200 kg are on the market, but only 10% of emer- gency units in the United States have such machines [52].

Approximately 2% of emergency patients in the US have a weight of more than 160 kg and one quarter of CT scans and two thirds of MRI machines are not able to image such patients. Another paper found that 2% of ultrasound studies, 0.4% of abdominal MRIs, and 0.1% of abdomi- nal CTs are “limited by the body habitus” in obese pa- tients, which means they are inadequate. This does not include patients who cannot be imaged by CT or MRI due to excessive body weight exceeding the weight-bearing capacity of the table [53]. In obese patients, CT has to be used with caution, because the dose of irradiation is high- er than in normal weight patients; an increase in weight of 40% will elevate the dose of irradiation by 100% [54].

Ultrasound (US)

US is a very useful and well-established tool for ab- dominal imaging. US is readily available, comparatively inexpensive, can be used at the bedside and is easily accepted by the patient. In comparison with CT and MRI it has the advantages of safety, low-cost, and availabil- ity and can be repeated as often as necessary. It provides valuable information regarding the abdominal solid vis- cera and is an excellent technique for real time depiction of the presence of fluid in the peritoneal or pleural spaces.

US is also a useful tool in other fields of pathology, such as soft tissue pathology, gynecology and obstetrics.

The pathology of thyroid, salivary glands, cervical lymph nodes and the eye can also be characterized by means of US. US can also serve in guiding therapeutic and diag- nostic procedures.

US has no absolute contraindications even if there ex- ist precautions to be taken in the first trimester of preg-

nancy: exposure time has to be limited at a minimum, mechanical and thermal indices have to be set as low as possible and Doppler ultrasound is to be avoided [55].

The use of US during pregnancy for non-medical purpos- es such as obtaining fetal videos has been discouraged.

1) Difficulties and sources of error in ultrasound As with other diagnostic methods, errors may occur in ultrasound examinations [56] due to multiple factors:

errors caused by the US machine, choice of transducers or setting (inadequate equipment quality); errors caused by the patient’s condition or preparation for the exam (e.g. in patients who cannot hold a breath for a couple of seconds, organs located behind the costal cage such as the liver or spleen may be difficult to examine; in patients who have eaten before ultrasound, it may not be possible to visualize the pancreas); failure to obtain an optimal image of the tested organs (in Chilaiditi syndrome parts of the liver may not be visualized due to interposition of the transverse colon; in some patients, parts of the pan- creas cannot be seen on ultrasound). Meteorism is a con- cern for abdominal ultrasound.

A retrospective study focusing on ultrasound of cir- rhotic livers for the detection of hepatocellular carcinoma (HCC) found that 20% of screening USs were of inad- equate quality [57]. Body mass index and male gender were the strongest predictive factors for inadequate im- aging, with the most common reasons being rib shadow- ing and inadequate beam penetration resulting in visu- alization of less than two thirds of the liver. A severely shrunken liver in advanced cirrhosis is difficult to de- scend below the rib cage.

2) US and obesity

US is the imaging technique most affected by obesity, because of the depth of penetration required with result- ant attenuation of the ultrasound beam [51]. In the imag- ing of an obese patient with 8-cm of subcutaneous fat, 94% of the original sound wave is attenuated before it reaches the peritoneal cavity [58]. A retrospective study of over 5 000 000 imaging studies found that 1.5% of abdominal US examinations were unsatisfactory due to obesity. This number is increasing with the increase in the prevalence of obesity. In 1989, only 0.7% of abdomi- nal ultrasounds were considered unsatisfactory due to obesity; this percent rose to 1.9% in 2003 [59].

There are some techniques available that can improve the visualization of peritoneal organs in obese patients, including the use of the lowest frequency of a multifre- quency transducer (the “penetration” mode) and har- monic imaging, which reduces the deleterious effects of subcutaneous tissues [60]. The signal-to-noise ratio can

(7)

Table I. Lists with factors which lead to inadequate imaging and solutions to obtain correct images in these patients Imaging

technique Factors which make the examination unrealizable

Factors which lead to

image degradation Percentage of

patients affected What to do in these circumstances

MRI Claustrophobia 1-2% Using open scanners

Sedation of patients Reducing the noise Presence of metallic

implants, devices or foreign bodies

~2.7 USA citizens Removal of the foreign body (if ferromag- netic)

Using MR conditional or MR safe devices Metallic implants or

devices which are not absolute contraindication to MRI

0.5-1% Using a high readout bandwidth Swapping frequency and phase Using thin sections

Lower magnet strength. Artifacts are more important at 3.0T as compared to 1.5T Increase the matrix

Increase the number of excitations NEX Shorter echo-spacing

Factors related to the patient (breathing or moving during the ex- amination)

Stabilization of the patient

Using surface coils confined to the area of interest

Using spatial saturation pulses Flow saturation pulses

Increasing number of signals averaged (NEX)

Swapping frequency and phase encoding directions

Use of single-slice ultrafast sequences Using self-correcting sequences like PRO- PELLER and BLADE

Factors related to the machine or the technique used

MR accounts for >90% recalls among imaging techniques

Good quality and maintenance of MRI machines.

Using highly-qualified MR technicians Contraindications to

contrast media adminis- tration

0.17-2.8% Premedication of patients; changing the type of contrast agent

Morbid obesity 2-3% in the USA Use of large weight capacity scanners

CT Limiting irradiation in patients where diag- nostic benefit of CT is low or when CT can be replaced with another imaging technique

Using correct and appropriate indica- tions will reduce the number of CT- scans by 3%

Correct choice of the imaging technique.

Using CT only when it is unavoidable Using low-dose protocols

Metallic implants or foreign bodies leading to image degradation

Remove metallic objects Usage of post- processing techniques such as Metal Dele- tion Techniques

Factors related to the patient (breathing or moving during the ex- amination)

<3% of cardiac CT exams are unin- terpretable due to cardiac motion

Using faster scanners Sedation of patients

Use of beta-blockers to reduce the heart frequency and variability (for cardiac CT).

ECG-editing Contraindications to

contrast media adminis- tration

0.2-0.6% of patients develop mild or moderate allergic reactions to contrast media;

Using an alternative imaging technique or premedication for patients with prior aller- gies to contrast media

Morbid obesity 2-3% in the USA Use of large weight capacity scanners

(8)

Ultrasound Uncooperative patients Sedation

Obesity >3% in the USA Reducing the frequency of the ultrasound beamUsing tissue harmonics

Bloating Dietary restrictions

Physical exercise Premedication MRI – magnetic resonance imaging; CT – computer tomography; USA – United States of America

also be improved by the use of compound imaging and the presence of speckle reduction filters [61]. In pelvic ultrasound, a transvaginal or transrectal approach can im- prove visualization.

3) Contrast-enhanced ultrasound (CEUS)

The main indication for CEUS is characterization of focal liver lesions (FLL). The quality of the image on CEUS depends on the quality of the image on US; when the patient is obese or bloated or the lesion is too pro- found, CEUS will be inadequate [62]. In a large multi- center study, CEUS had sufficient diagnostic quality to allow assessment of 1280 out of 1349 focal liver lesions (94.9%). The remaining 5.1% of FLL could not be appro- priately characterized because of adiposity of the patient, steatosis of the liver, the presence of calcifications or air interference [63].

General considerations

Even if imaging has shown substantial progress over the last few decades, diagnostic limitations still exists for each technique. With 33% of American adults being obese, a percentage is increasing, obesity is a major pub- lic health hazard and is influencing the ability to image pathology in these patients [60]. Obese patients present limitations for the use and diagnostic capabilities of US, CT and MRI. Furthermore, 3% of American adults are considered morbidly obese, and imaging them can be extremely difficult, sometimes only being possible in centers with dedicated CT and MRI machines [64]. Also, obese patients tend to show greater mortality from cancer than normal weight patients. This may be due to difficul- ties in both diagnosis and treatment of cancer in obese patients [65].

The imaging of patients with chronic renal disease and a GFR lower than 30 ml/min/1.73 m2 is challenging, as administration of either iodinated contrast media or gadolinium products is a relative contraindication. Ad- ministration of Sonovue is not contraindicated in these patients, and CEUS can represent an alternative diagnos- tic tool; for example, in characterization of FLL where

contrast injection is mandatory. Furthermore, recent FDA regulations have requested that only Omniscan, Magne- vist and Optimark should be contraindicated in the set- ting of renal insufficiency. If contrast administration is absolutely necessary to obtain a correct diagnosis, the radiologist should inject one of the other GdCA available on the market (Ablavar, Dotarem, Eovist, Gadavist, Mul- tiHance or Prohance) at the lowest sufficient dose [66].

MRI is an imaging technique performed in an un- friendly environment due to the closed scanner bore and the presence of acoustic noise. This environment leads to increased anxiety, particularly in patients with head injuries, psychiatric disorders, the elderly and infants [67]. However, anxiety leading to premature termination of examinations and consequently to inadequate imag- ing, can affect patients regardless of their age, gender or preexisting psychiatric pathology. Recent studies have demonstrated decreases in both claustrophobic symp- toms and premature termination of exams with the use of modern MRI scanners with a shorter and wider bore and reduced acoustic noise (from 3-5% to 1-2%). However, many studies underestimate the number of patients un- able to perform MRI, because they do not include those patients who completely refuse it.

Motion artifacts (e.g. in uncooperative patients) can lead to image degradation and to uninterpretable exami- nations in both CT and MRI. If patients cannot be sedat- ed, then CT could be a better option because it is quicker to perform and less prone to artifacts.

Metallic artifacts can affect both techniques, with MR examinations once again being more influenced, and sometimes being rendered uninterpretable. Modalities for reducing the severity of metallic artifacts are nowa- days available for both techniques.

The impossibility of MR imaging for patients with MR unsafe implants and devices is a major public health issue. Most of the implants and devices manufactured nowadays are either “MRI-safe” or “MRI-conditional”.

Table I lists the factors which lead to inadequate im- aging and solutions to obtain correct images.

Conflict of interest: none

(9)

References

1. Dean Deyle G. The role of MRI in musculoskeletal prac- tice: a clinical perspective. J Man Manip Ther 2011;19:152- 2. Maniam S, Szklaruk J. Magnetic resonance imaging: Re-161.

view of imaging techniques and overview of liver imaging.

World J Radiol 2010;2:309-322.

3. Kennedy AM, Gilfeather MR, Woodward PJ. MRI of the female pelvis. Semin Ultrasound CT MR 1999;20:214- 4. Desai MJ, Hargens LM, Breitenfeldt MD, et al. The rate 230.

of magnetic resonance imaging in patients with spinal cord stimulation. Spine (Phila Pa 1976) 2015;40:E531-E537.

5. Enders J, Zimmermann E, Rief M, et al. Reduction of claus- trophobia during magnetic resonance imaging: methods and design of the “CLAUSTRO” randomized controlled trial. BMC Med Imaging 2011;11:4.

6. ASTM F2503-05: Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Res- onance Environment, ASTM International, West Consho- hocken, PA, 2005, www.astm.org.

7. Association of Anaesthetists of Great Britain and Ireland, Farling PA, Flynn PA, et al. Safety in magnetic resonance units: an update. Anaesthesia 2010;65:766-770.

8. Expert Panel on MR Safety, Kanal E, Barkovich AJ, et al.

ACR guidance document on MR safe practices: 2013. J Magn Reson Imaging 2013;37:501-530.

9. Shellock FG.. Reference Manual for Magnetic Resonance Safety, Implants, and Devices. Los Angeles, CA: Biomedi- cal Research Publishing Group, 2013.

10. Russo RJ, Costa HS, Silva PD, et al. Assessing the Risks Associated with MRI in Patients with a Pacemaker or De- fibrillator. N Engl J Med 2017;376:755-764.

11. Stacul F, van der Molen AJ, Reimer P, et al. Contrast in- duced nephropathy: updated ESUR Contrast Media Safety Committee guidelines. Eur Radiol 2011;21:2527-2541.

12. Quirk M, Ciottone R, Letendre D, Wapner S. Critical per- son-in-environmental transitions in medical education.

Med Teach 1987;9:415-423.

13. Datendorfer K, Amering M, Bankier A, et al. A study of the effect of patient anxiety, perception and equipment on mo- tion artifacts in magnetic resonance imaging. Magn Reson Imaging 1997;15:301-306.

14. Melendez JC, McCrank E. Anxiety-related reactions as- sociated with magnetic resonance imaging examinations.

JAMA 1993;270:745-747.

15. Berg WA, Blume J, Adams AM, et al. Reasons women at elevated risk of breast cancer refuse breast MR imaging screening: ACRIN 6666. Radiology 2010;254:79–87.

16. Dewey M, Schink T, Dewey CF. Claustrophobia during magnetic resonance imaging: cohort study in over 55,000 patients. J Magn Reson Imaging 2007;26:1322-1327.

17. Bangard C, Paszek J, Berg F, et al. MR imaging of claus- trophobic patients in an open 1.0T scanner: motion artifacts and patient acceptability compared with closed bore mag- nets. Eur J Radiol 2007;64:152-157.

18. Koch KM, Lorbiecki JE, Hinks RS, King KF. A multispec- tral three-dimensional acquisition technique for imaging near metal implants. Magn Reson Med 2009;61:381-390.

19. Lu W, Pauly KB, Gold GE, Pauly JM, Hargreaves BA. SE- MAC: slice encoding for metal artifact correction in MRI.

Magn Reson Med 2009:66-76.

20. Firmin D, Keegan J. Navigator echoes in cardiac magnetic resonance. J Cardiovasc Magn Reson 2001;3:183-193.

21. Lee VS, Lavelle MT, Rofsky NM, et al. Hepatic MR imag- ing with a dynamic contrast-enhanced isotropic volumetric interpolated breath-hold examination: feasibility, reproduc- ibility, and technical quality. Radiology 2000;215:365-372.

22. Zaitsev M, Maclaren J, Herbst M. Motion artifacts in MRI:

a complex problem with many partial solutions. J Magn Reson Imaging 2015;42:887-901.

23. Jung JW, Kang HR, Kim MH, et al. Immediate hypersen- sitivity reaction to gadolinium-based MR contrast media.

Radiology 2012;264:414-422.

24. Murphy KJ, Brunberg JA, Cohan RH. Adverse reactions to Gadolinium contrast media- a review of 36 cases. Am J Roentgenol 1996;167:847-849.

25. Li A, Wong CS, Wong MK, Lee CM, Au Yeung MC. Acute adverse reactions to magnetic resonance contrast me- dia--gadolinium chelates. Br J Radiol 2006;79:368-371.

26. Becker S, Walter S, Witzke O, Kreuter A, Kribben A, Mitch- ell A. Application of gadolinium-based contrast agents and prevalence of nephrogenic systemic fibrosis in a cohort of end-stage renal disease patients on hemodialysis. Nephron Clin Pract 2012;121:c91-c94.

27. Lima XT, Alora-Palli MB, Kimball AB, Kay J. Validation of a screening instrument for nephrogenic systemic fibrosis.

Arthritis Care Res (Hoboken) 2013;65:637-642.

28. Sosna J, Slasky BS, Bar-Ziv J. Computed tomography in the emergency department. Am J Emerg Med 1997;15:244-247.

29. Tsapaki V, Rehani M, Saini S. Radiation safety in abdom- inal computed tomography. Semin Ultrasound CT MR 2010;31:29-38.

30. Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M. Effec- tive doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 2008;248:254-263.

31. National Council on Radiation Protection and Measure- ments. NCRP Report No. 160, Ionizing radiation exposure of the population of the United States. 2009.

32. Boedeker KL, McNitt-Gray MF. Application of the noise power spectrum in modern diagnostic MDCT: part II.

Noise power spectra and signal to noise. Phys Med Biol 2007;52:4047-4061.

33. Huda W, Ogden KM, Samei E, et al. Reconstruction filters and contrast detail curves in CT. Proc. SPIE 6917, Medical Imaging 2008: Image Perception, Observer Performance, and Technology Assessment, 691710 (6 March 2008); doi:

10.1117/12.770530.

34. Mahesh M, Durand DJ. The choosing wisely campaign and its potential impact on diagnostic radiation burden. J Am Coll Radiol 2013;10:65-66.

35. Sistrom CL, Dang PA, Weilburg JB, Dreyer KJ, Rosenthal DI, Thrall JH. Effect of computerized order entry with inte-

(10)

grated decision support on the growth of outpatient proce- dure volumes: seven-year time series analysis. Radiology 2009;251:147-155.

36. Morris KT, Kavanagh M, Hansen P, Whiteford MH, De- veney K, Standage B. The rational use of computed tomog- raphy scans in the diagnosis of appendicitis. Am J Surg 2002;183:547-550.

37. Barrett JF, Keat N. Artifacts in CT: recognition and avoid- ance. Radiographics 2004;24:1679-1691.

38. Boas FE, Fleischmann D. Evaluation of two iterative tech- niques for reducing metal artifacts in computed tomogra- phy. Radiology 2011;259:894-902.

39. Andersson KM, Nowik P, Persliden J, Thunberg P, Norrman E. Metal artefact reduction in CT imaging of hip prostheses - an evaluation of commercial techniques provided by four vendors. Br J Radiol 2015;88:20140473.

40. Fleischmann D, Boas FE. Computed tomography – old ide- as and new technology. Eur Radiol 2011;21:510-517.

41. Prat-Gonzalez S, Sanz J, Garcia MJ. Cardiac CT: Indica- tions and Limitations. Nucl Med Technol 2008;36:18-24.

42. Kalisz K, Buethe J, Saboo SS, Abbara S, Halliburton S, Rajiah P. Artifacts at Cardiac CT: Physics and Solutions.

Radiographics 2016;36:2064-2083.

43. Oliver JH 3rd, Baron RL. Helical biphasic contrast-en- hanced CT of the liver: technique, indications, interpreta- tion, and pitfalls. Radiology 1996;201:1-14.

44. Wang CL, Cohan RH, Ellis JH, Caoili EM, Wang G, Fran- cis IR. Frequency, outcome, and appropriateness of treat- ment of nonionic iodinated contrast media reactions. AJR Am J Roentgenol 2008;191:409-415.

45. Cochran ST, Bomyea K, Sayre JW. Trends in adverse events after IV administration of contrast media. AJR Am J Roentgenol 2001;176:1385-1388.

46. Park HJ, Park JW, Yang MS, et al. Re-exposure to low os- molar iodinated contrast media in patients with prior mod- erate-to-severe hypersensitivity reactions: A multicentre retrospective cohort study. Eur Radiol 2017;27:2886-2893.

47. Katayama H, Yamaguchi K, Kozuka T, Takashima T, Seez P, Matsuura K. Adverse reactions to ionic and nonionic con- trast media. A report from the Japanese Committee on the Safety of Contrast Media. Radiology 1990;175:621-628.

48. Yamaguchi K, Katayama H, Takashima T, Kozuka T, Seez P, Matsuura K. Prediction of severe adverse reactions to ionic and nonionic contrast media in Japan: evaluation of pretesting. A report from the Japanese Committee on the Safety of Contrast Media. Radiology 1991;178:363-367.

49. Fischer HW, Doust VL. An evaluation of pretesting in the problem of serious and fatal reactions to excretory urogra- phy. Radiology 1972;103:497-501.

50. Abe S, Fukuda H, Tobe K, Ibukuro K. Protective effect against repeat adverse reactions to iodinated contrast medi- um: Premedication vs. changing the contrast medium. Eur Radiol 2016;26:2148-2154.

51. Glanc P, O’Hayon BE, Singh DK, Bokhari SA, Maxwell CV. Challenges of pelvic imaging in obese women. Radio- graphics 2012;32:1839-1862.

52. Ginde AA, Foianini A, Renner DM, Valley M, Camargo CA Jr. The challenge of CT and MRI imaging of obese individ- uals who present to the emergency department: a national survey. Obesity (Silver Spring) 2008;16:2549-2551.

53. Miller JC, Phil D. Imaging and Obese Patients. Radiology Rounds 2005;3:1-4.

54. Israel GM, Cicchiello L, Brink J, Huda W. Patient size and radiation exposure in thoracic, pelvic, and abdominal CT examinations performed with automatic exposure control.

AJR Am J Roentgenol 2010;195:1342-1346.

55. Ter Haar G. Ultrasonic imaging: safety considerations. In- terface Focus 2011;1:686-697.

56. Walas MK, Skoczylas K, Gierblinski I. Errors and mistakes in the ultrasound diagnostics of the liver, gallbladder and bile ducts. J Ultrason 2012;12:446-462.

57. Simmons O, Fetzer DT, Yokoo T, et al. Predictors of ad- equate ultrasound quality for hepatocellular carcinoma surveillance in patients with cirrhosis. Aliment Pharmacol Ther 2017;45:169-177.

58. Bushberg JT, Seibert JA, Leidholdt EM, Boone JM. The essential physics of medical imaging. 2nd ed. Philadelphia:

Lippincott Williams & Wilkins, 2002.

59. Uppot RN, Sahani DV, Hahn PF, Kalra MK, Saini SS, Mueller PR. Effect of obesity on image quality: fifteen-year longitudinal study for evaluation of dictated radiology re- ports. Radiology 2006;240:435-439.

60. Modica MJ, Kanal KM, Gunn ML. The Obese Emergency Patient: Imaging Challenges and Solutions. Radiographics 2011;31:811-823.

61. Paladini D. Sonography in obese and overweight pregnant women: clinical, medicolegal and technical issues. Ul- trasound Obstet Gynecol 2009;33:720-729.

62. Chiorean L, Tana C, Braden B, et al. Advantages and Lim- itations of Focal Liver Lesion Assessment with Ultrasound Contrast Agents: Comments on the European Federation of Societies for Ultrasound in Medicine and Biology (EF- SUMB) Guidelines. Med Princ Pract 2016;25:399-407.

63. Strobel D, Seitz K, Blank W, et al. Contrast-enhanced ul- trasound for the characterization of focal liver lesions--di- agnostic accuracy in clinical practice (DEGUM multicenter trial). Ultraschall Med 2008;29:499-505.

64. Sturm R, Hattori A. Morbid obesity rates continue to rise rapidly in the United States. Int J Obes (Lond) 2013;37:889- 65. Calle EE, Rodriguez C, Walker-Thurmond K, Thun MJ. 891.

Overweight, obesity, and mortality from cancer in a pro- spectively studied cohort of U.S. adults. N Engl J Med 2003;348:1625-1638.

66. FDA U.S. Food and Drug Administration. FDA Drug Safety Communication: New warnings for using gado- linium-based contrast agents in patients with kidney dys- function. Available at: www.fda.gov/Drugs/Drugsafety/

ucm223966

67. Quirk ME, Letendre AJ, Ciottone RA, Lingley JF. Anx- iety in patients underogoing MR imaging. Radiology 1989;170:463-466.

Referințe

DOCUMENTE SIMILARE

10 As a matter of fact, this alternative (or: path in the specialisation search space) can be pruned by analysing the parses eventually provided as annotation to the

Accordingly, GI tract obstruction was ruled out correctly in 2 patients and diagnosed correctly in 12 patients, in- cluding 7 patients with Hirschsprung’s disease and 5 pa- tients

Faced with the possible insurrection of the body and of the sensible in general, and in order to surpass the possible nefarious consequences of such a logical dead end, (clear

For example, an entity that is classified as an association or a sports club under national law may nevertheless have to be regarded as an undertaking within the meaning of

According to our previous investigations it seems that tolerance, whether regarded as a political practice or a philosophical or moral principle, is a strategy (or tactics) of one