Evaluation of the effect of computer aided diagnosis system

on breast ultrasound for inexperienced radiologists in describing and determining breast lesions

Jeongmin Lee, Sanghee Kim, Bong Joo Kang, Sung Hun Kim, Ga Eun Park

Departments of Radiology, Seoul St. Mary’s Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea

Received 20.02.2019 Accepted 04.06.2019 Med Ultrason

2019, Vol. 21, No 3, 239-245

Corresponding author: Bong Joo Kang, MD, PhD, Professor Department of Radiology, Seoul St. Mary’s Hospital, College of Medicine,

The Catholic University of Korea 222, Banpo-daero, Seocho-gu, Seoul, 06591, Republic of Korea

Phone: 82-2-2258-6253 Fax: 82-2-599-6771

E-mail: [email protected]

Introduction

In the development of medical technology, new ad- vanced methods are emerging that actively assist in the diagnosis and treatment. The most representative method in the radiological field is the computer-aided detec- tion or diagnosis (CAD) system. The first CAD system commercially approved by the Food and Drug Admin- istration (FDA) in 1998, was to automatically detect and mark calcifications or masses in digital mammography Abstract

Aim: To investigate the effect of a computer-aided diagnosis (CAD) system on breast ultrasound (US) for inexperienced radiologists in describing and determining breast lesions. Materials and methods: Between October 2015 to January 2017, 500 suspicious or probable benign lesions in 413 patients were reviewed. Five experienced readers retrospectively reviewed for each of 100 lesions according to the Breast Imaging Reporting and Data System (BI-RADS) lexicon and category, with CAD system (S-detectTM). The readers then made final decisions by combining CAD results to their US results. Using the nested experiment design, five inexperienced readers were asked to select the appropriate BI-RADS lexicons, categories, CAD results, and combination results for each of the 100 lesions, retrospectively. Diagnostic performance of experienced and inexperienced radiologists and CAD were assessed. For each case, agreements in the lexicons and categories were analyzed among the experienced reader, inexperienced reader and CAD. Results: Indicators of the diagnostic performance for breast malignancy of the experienced group (AUC=0.83, 95%CI [0.80, 0.86]) were similar or higher than those of CAD (AUC = 0.79, 95%CI[0.74, 0.83], p=0.101), except for specificity. Conversely, indicators of diagnostic performance of inexperienced group (AUC=0.65, 95%CI[0.58, 0.71]) did not differ from or were lower than those of CAD(AUC=0.73, 95%CI[0.67, 0.78], p=0.013). Also, the diagnostic performance of the inexperienced group after combination with the CAD result was signifi- cantly improved (0.71, 95% CI [0.65, 0.77], p=0.001), whereas that of the experienced group did not change after combination with the CAD result, except for specificity and positive predictive value (PPV). Kappa values for the agreement of the catego- rization between CAD and each radiologist group were increased after applying the CAD result to their result of general US.

Especially, the increase of the Kappa value was higher in the inexperienced group than in the experienced group. Also, for all the lexicons, the Kappa values between the experienced group and CAD were higher than those between the inexperienced group and CAD. Conclusion: By using the CAD system for classification of breast lesions, diagnostic performance of the inexperienced radiologists for malignancy was significantly improved, and better agreement was observed in lexicons between the experienced group and CAD than between the inexperienced group and CAD. CAD may be beneficial and educational for the inexperienced group.

Keywords: ultrasound; computer-aided diagnosis; breast cancer

DOI: 10.11152/mu-1889

(2)

[1]. While several previous studies have suggested that CAD systems can contribute to increase the early cancer detection rates, the number of breast radiologists actively relying on this CAD system for mammography readings is small [2].

In addition to simplify the lesions’ detection, the CAD systems for breast lesion have been introduced in magnetic resonance imaging (MRI) and ultrasound (US) techniques. MRI CAD system automatically analyzes the kinetics of enhancing lesions on breast MRI and it can contribute to improve the specificity and reduce the in- terpretation time by radiologists [1,3-5].

Breast US is one of the commonly used screening methods in clinical practice as a supplementary tool es- pecially in women with dense breast tissue [6]. However, US is more operator dependent than other methods. For this reason, many previous studies have demonstrated the efficiency and usefulness of CAD systems for improving diagnostic accuracy in US [7-12]. In our previous study, a CAD system newly developed in Korea (S-Detect, Samsung Medison Co. Ltd., Seoul, Korea) was useful for less experienced radiologists and the combination of the CAD system and general US improved the specificity in inexperienced and experienced radiologists [13].

The purpose of this study was to investigate the effect of CAD (S-detectTM) on breast US when inexperienced radiologists described and categorized breast lesions, es- pecially in comparison with experienced radiologists.

Materials and methods

This retrospective study was approved by our Institu- tional Review Board (IRB) and all enrolled participants provided written informed consent. A total of 682 breast lesions from 566 patients examined by US between Oc- tober 2015-January 2017 were consecutively obtained using the Samsung Ultrasound RS80A, Samsung Medi- son Co. Ltd., Seoul, Korea machine by five experts in breast radiologists, with more than 5 years’ experience.

The cases without pathologic confirmation (N = 19) and findings associated with post-operative changes (N = 163) such as post-operative fibrosis or fat necrosis, were excluded. Finally, the data of 500 suspicious or probable benign breast lesions from 413 patients were enrolled.

All the lesions were examined using the CAD system (S-detectTM) at the same time as general US. When ra- diologists identified the center of the breast lesion by touching the screen, a region of interest (ROI) was au- tomatically drawn along the border of the lesion. Several drawn borders were presented on the screen and the ra- diologists saved more than two CAD images as with the general US.

Imaging analysis

All 500 cases were retrospectively analyzed by two radiologist groups. The first radiologist group comprised five experienced radiologists with clinical fellowship in breast imaging and more than five years’ experience of breast US. Expert breast radiologists who performed US and extracted cases, were excluded from this group. The second radiologist group was composed with five inexpe- rienced radiologists with less than one year of breast US.

Firstly, experienced and inexperienced groups de- scribed and categorized the lesions according to the Breast Imaging Reporting and Data System (BI-RADS) lexicon: shape, orientation, margin, echogenicity and acoustic shadowing. In the first conclusions using gener- al US, description and categorization of lesions by CAD (S-detectTM) were disclosed to two radiologist groups.

The CAD system described the lesions according to the BI-RADS lexicons and they were classified in to two categories: “possibly benign” or “possibly malignant”.

Readers selected the most appropriate CAD image as with the general US. After referring to results by CAD, experienced and inexperienced groups were given the opportunity to re-categorize the lesions and they made final subjectively combined conclusion for the lesions (fig 1). The diagnostic performance of both radiologist groups and CAD were compared, and the gold standard of diagnosis was the pathologic report presented after surgery or the follow-up breast US realized more than one year later.

Fig 1. Flow chart of exclusion and decision-making for breast lesions detected from screening or diagnostic US by both expe- rienced and inexperienced radiologist groups.

(3)

Statistical analysis

The diagnostic performance for breast malignancy including the area under the receiver operating charac- teristic (ROC) curve (AUC), sensitivity and specificity of the two radiologist groups and CAD were calculated.

The same parameters were then also calculated for the combination conclusions of each group combining the gray-scale US results and the CAD results.

After evaluation of the diagnostic performance of CAD and each radiologist group, we evaluated the agreement of BI-RADS US lexicon and final categori- zation between CAD and each radiologist group using Kappa statistics. The estimated Kappa used to estimate the extent of agreement was defined by Landis et al [14]

as follows: if Kappa is less than 0, “No agreement”, if 0-0.2, “Slight agreement”, if 0.2-0.4, “Fair agreement”, if 0.4-0.6, “Moderate agreement”, if 0.6-0.8, “Substantial agreement”, and if 0.8-1.0, “Almost perfect agreement”.

The statistical analysis was performed using the Sta- tistical Package for the Social Sciences (SPSS) Statistics software 24.0 (IBM Corp, Armonk, New York). A p val- ue less than 0.05 was statistically significant.

Result

The demographic characteristics of the 413 patients with 500 lesions are detailed in table I.

Diagnostic performance of radiologists and CAD Firstly, comparing the results of general US obtained by experienced radiologists with those of inexperienced radiologists, all diagnostic performance indicators were significantly higher in the experienced group (p<0.001), except for specificity, that did not differ between the two groups (p=0.798).

The diagnostic accuracy for breast malignancy of the experienced radiologist group was high (AUC=0.83, 95% CI [0.80, 0.86]) and no significant difference was

observed with the CAD result (AUC=0.79, 95%CI [0.74, 0.83], p=0.101). However, the diagnostic accuracy of the inexperienced group was 0.65 (95%CI [0.58, 0.71]), lower than that of the CAD result (AUC=0.73, 95%CI [0.67, 0.78], p=0.013).

The AUC value of the combination conclusion (gen- eral US and CAD) of the experienced group was not significantly different from the AUC value of the gen- eral US only result by the experienced group (Table II).

However, in the inexperienced group, the combination conclusion revealed higher AUC value than the general US only result. Not only the AUC value of the combina- tion conclusion in the inexperienced group, but other fac- tors including sensitivity, specificity, PPV and negative predictive value (NPV) were higher. This means that the inexperienced group made a more accurate judgment for breast lesions when making the conclusion based on the CAD results (fig 2).

No significant differences were revealed in other diagnostic performance indicators of the combination conclusion in the experienced group, compared to the results of general US, except for specificity and PPV. In the experienced group, sensitivity was higher than that of CAD, but CAD revealed higher specificity than specific- ity of general US in the experienced group.

Agreement of BI-RADS categorization and US lexicon between CAD and radiologists

We compared the categorization by CAD and radi- ologists for the suspicious or probable benign lesions on US. According to the BI-RADS category, lesions with more than C4A – low suspicion for malignancy – were re-categorized as “suspicious lesions” and the others as

“not suspicious lesions” by radiologists.

When comparing the categorizations, the agreement between CAD and general US of the experienced group reveals fair agreement with the Kappa value of 0.38 (95%CI [0.29, 0.47], p<0.001) and moderate agreement

Table I. Patients’ characteristics

Total Benign Malignancy p value

Patients 413 354(85.74) 59(14.29)

Lesions 500 432(86.4) 68(13.6)

Age (years) 48.76±11.53 48.13±11.64 52.53±10.18 0.026

US for screening 272(65.9) 243(89.34) 29(10.66) 0.006

US for diagnostic 141(34.1) 111(78.72) 30(21.28)

Symptoms’ presence 63(15.3) 36(57.14) 27(42.86)

Family history 109(26.4) 96(88.07) 13(11.93)

Postmenopausal status 177(42.9) 148(83.62) 29(16.38)

Size (cm) 1.19±0.78 1.15±0.77 1.46±0.76 <0.001

The results are expressed as number, mean±standard deviation or mean (%). US – ultrasonography.

(4)

Table II. Diagnostic performance of combination conclusion by experienced and inexperienced radiologist groups AUCp valueSensitivityp valueSpecificityp valuePPVp valueNPVp value Experienced group General ultrasound 0.83 (0.8-0.86)

reference

0.97 (0.90-1.00)

reference

0.7 (0.65-0.74)

reference

0.34 (0.27-0.41)

reference

0.99 (0.98-1.00)

reference S-detect

0.79 (0.74-0.84)

0.101

0.81 (0.70-0.89)

<0.001

0.77 (0.73-0.81)

0.004

0.36 (0.28-0.44)

0.351

0.96 (0.94-0.98)

0.001 Subjective combined conclusion

0.84 (0.81-0.87)

0.451

0.96 (0.88-0.99)

0.317

0.72 (0.68-0.77)

0.003

0.35 (0.28-0.43)

0.025

0.99 (0.97-1.00)

0.358 Inexperienced group General ultrasound

0.65 (0.58-0.71)

reference

0.59 (0.46-0.71)

reference

0.7 (0.66-0.75)

reference

0.24 (0.18-0.31)

reference

0.92 (0.88-0.94)

reference S-detect

0.73 (0.67-0.78)

0.013

0.76 (0.65-0.86)

0.007

0.69 (0.65-0.74)

0.710

0.28 (0.22-0.35)

0.069

0.95 (0.92-0.97)

0.009 Subjective combined conclusion

0.71 (0.65-0.77)

0.001

0.69 (0.57-0.8)

0.008

0.73 (0.69-0.77)

0.033

0.29 (0.22-0.37)

< 0.001

0.94 (0.91-0.96)

0.003 Table III.Kappa values between CAD and radiologist group for determination (final category) and descriptions (lexicon) Kappa statistics Experienced groupInexperienced group DeterminationFinal Category0.38 (0.29-0.47)0.4 (0.32-0.49) Combined Conclusion†0.44 (0.53-0.36)0.57 (0.50-0.65) Descriptions (Lexicon)Shape 0.51 (0.43-0.58)0.3 (0.22-0.38) Orientation0.58 (0.45-0.71)0.29 (0.16-0.42) Margin0.37 (0.3-0.44)0.25 (0.18-0.32) Echo pattern0.43 (0.36-0.51)0.12 (0.04-0.2) Posterior features0.37 (0.27-0.47)0.22 (0.11-0.33) † “Combined Conclusion” means subjective combined conclusion that appropriate decision was chosen subjectively, based on final US category with CAD conclusion.

(5)

with Kappa value 0.4 (95%CI [0.32, 0.49]) between CAD result and general US of inexperienced group was noted. However, after applying the CAD result to the general US of each group, increases of Kappa value were revealed in both groups, and were more prominent in the inexperienced group (from 0.38 to 0.44 in experienced group, from 0.4 to 0.57 in inexperienced group).

The Kappa values for all US lexicons between CAD and each radiologist group were also evaluated. The Kap- pa value between CAD and experienced group is higher than that between CAD and inexperienced group in all US lexicons (Table III, fig 3).

Discussions

Breast cancer is the most frequently diagnosed can- cer in women worldwide. However, due to its favorable prognosis it ranks as the fifth cause of deaths in Korea

[15]. For this reason, early detection and diagnosis of breast cancer is critical for survival [16-18]. Mammogra- phy is the only screening method proven to improve early diagnosis and decrease morbidity and mortality of breast cancer. However, in dense breast which is classified as high-density pattern – C or D – in the BI-RADS lexicon, the sensitivity of mammography to detect cancer is poor.

Therefore, breast US was adopted as a complementary method as a screening tool for dense breasts [19].

Breast US is potentially effective for screening, but due to its operator-dependent nature, many additional US techniques have recently been suggested and applied in practice to improve diagnostic accuracy. One of the newly suggested techniques is the CAD system, that can provide efficient interpretation or second opinion for a lesion detected on breast US [10,12].

S-detectTM (Samsung Medison, Co, Ltd, Seoul, Ko- rea) is a recently developed US CAD system that pro-

Fig 2. A hypoechoic mass was detected at 10 o’clock direction of left breast on general US (A) for screening of 43-year-old woman.

For this lesion, CAD analysis was applied (B) at the same time of general US. An experienced radiologist categorized this lesion as BI-RADS category 4C (high suspicion for malignancy), and the inexperienced radiologist categorized it as BI-RADS category 3 (probably benign lesion). From the CAD result, both groups finally categorized the lesion as a suspicious lesion of BI-RADS cat- egory 4 or higher. This lesion was confirmed as an invasive ductal carcinoma after excision.

Fig 3. A hypoechoic mass was detected at 9 o’clock direction of the right breast on general US (A) for a diagnostic sognography of 47-year-old woman with palpable mass. Both experienced and inexperienced radiologists described and categorized the mass ac- cording to the BI-RADS lexicon. The CAD result (B) was compared with both radiologists’ results repectively. While no differences were found in the description between CAD and the experienced radiologist, some differences were found between the CAD result and the inexperienced radiologist’s result. Categorization of the mass by radiologists and CAD was the same. After excision, this lesion was confirmed as an invasive ductal carcinoma.

(6)

vides additional morphologic analysis of breast masses detected on breast US according to the BI-RADS lexi- con and assists in the final assessment for the detected breast masses [20]. According to previous studies using S-detectTM, it can be beneficial to improve specificity for the diagnosis of malignant breast mass in breast US regardless of the experience of radiologists [13,21]. In particular, the use of S-detectTM showed increased speci- ficity of the less experienced radiologists compared to the experienced breast radiologists in previous studies done by our group [13] and by Di Segni et al [22].

Similar to previous studies, application of the CAD result to the result of general US of each radiologist group showed significant increase of specificity of both groups in our study. In the inexperienced group, not only speci- ficity, but other indicators of diagnostic performance sig- nificantly increased. However, unlike previous studies on S-detectTM, this study has a significantly greater number of lesions analyzed (n=500), and each group was com- posed of similarly experienced radiologists.

The Kappa statistics showed that the agreement be- tween CAD and each radiologist group for categorization of suspicious lesions was increased after each radiologist group applied CAD result to their final categorization.

Moreover, increase of Kappa value was higher in the in- experienced group than in the experienced group. This implies that the inexperienced group is more likely to ac- cept the CAD result than the experienced group when making a final decision for the suspicious lesions on gen- eral US. Furthermore, not only the increase of Kappa val- ue but also the absolute Kappa value of the inexperienced group was higher than that of the experienced group after applying the CAD result.

Given these results, the US CAD system can serve as a supplementary tool for inexperienced groups when making a final decision about the status of a lesion de- tected on US. This also means that it can compensate for the disadvantages of operator-dependency of US.

Conversely, for the experienced group, CAD serves as a means of providing a second opinion similar to mam- mography CAD.

Agreement between CAD and each radiologist group for US lexicon was also reviewed in our study. In every lexicon, higher Kappa values were noted between CAD and the experienced group, than CAD and the inexperi- enced group. Although the result of CAD for US lexicon is not a gold standard, it may be beneficial and educa- tional for inexperienced radiologists to describe and un- derstand the features of a lesion with a CAD result.

Our study has a few limitations. First, due to its ret- rospective nature, analysis of the features of lesions was limited, because the radiologists did not analyze real-

time images, but captured images. Second, calcifications and non-mass lesions were not included in the analysis due to a lack of detection in general US. This situation can imply different results when applying CAD in clini- cal practice, because some cases of breast cancers appear as calcifications or non-mass lesions only.

We comparatively investigated the diagnostic perfor- mances of an experienced radiologist group and an in- experienced radiologist group to distinguish malignant and benign lesions when applying the CAD system in US. Overall, while no significant changes were observed in the experienced group when using the CAD system, a significant increase of diagnostic performance was re- vealed in the inexperienced group. Also, we found that the inexperienced group is more likely to accept and ap- ply the CAD result when making final decision for the categorization of the lesions through Kappa values be- tween CAD and each radiologist group.

In conclusion, the CAD system can be more ben- eficial and educational for less experienced radiologists than for experienced radiologists, not only when describ- ing lesions, but also when determining if the lesion is malignant.

Acknowledgement: The statistical consultation was supported by Department of Biostatistics of the Catholic Research Coordinating Center, Catholic Medical Center, South Korea.

Conflict of interest: none

References

1. Dromain C, Boyer B, Ferre R, Canale S, Delaloge S, Bal- leyguier C. Computed-aided diagnosis (CAD) in the detec- tion of breast cancer. Eur J Radiol 2013;82:417-423.

2. Freer TW, Ulissey MJ. Screening mammography with com- puter-aided detection: prospective study of 12,860 patients in a community breast center. Radiology 2001;220:781- 3. Wood C. Computer Aided Detection (CAD) for breast 786.

MRI. Technol Cancer Res Treat 2005;4:49-53.

4. Meeuwis C, van de Ven SM, Stapper G, et al. Computer- aided detection (CAD) for breast MRI: evaluation of ef- ficacy at 3.0 T. Eur Radiol 2010;20:522-528.

5. Meinel LA, Stolpen AH, Berbaum KS, Fajardo LL, Rein- hardt JM. Breast MRI lesion classification: improved per- formance of human readers with a backpropagation neural network computer-aided diagnosis (CAD) system. J Magn Reson Imaging 2007;25:89-95.

6. Berg WA, Blume JD, Cormack JB, et al. Combined screen- ing with ultrasound and mammography vs mammography alone in women at elevated risk of breast cancer. JAMA 2008;299:2151-2163.

(7)

7. Chen CM, Chou YH, Han KC, et al. Breast lesions on sono- grams: computer-aided diagnosis with nearly setting-inde- pendent features and artificial neural networks. Radiology 2003;226:504-514.

8. Kim KG, Cho SW, Min SJ, Kim JH, Min BG, Bae KT.

Computerized scheme for assessing ultrasonographic fea- tures of breast masses. Acad Radiol 2005;12:58-66.

9. Drukker K, Giger ML, Vyborny CJ, Mendelson EB. Com- puterized detection and classification of cancer on breast ultrasound. Acad Radiol 2004;11:526-535.

10. Horsch K, Giger ML, Vyborny CJ, Venta LA. Performance of computer-aided diagnosis in the interpretation of lesions on breast sonography. Acad Radiol 2004;11:272-280.

11. Sehgal CM, Cary TW, Kangas SA, et al. Computer- based margin analysis of breast sonography for differen- tiating malignant and benign masses. J Ultrasound Med 2004;23:1201-1209.

12. Horsch K, Giger ML, Vyborny CJ, Lan L, Mendelson EB, Hendrick RE. Classification of breast lesions with multimo- dality computer-aided diagnosis: observer study results on an independent clinical data set. Radiology 2006;240:357-368.

13. Choi JH, Kang BJ, Baek JE, Lee HS, Kim SH. Application of computer-aided diagnosis in breast ultrasound interpreta- tion: improvements in diagnostic performance according to reader experience. Ultrasonography 2018;37:217-225.

14. Landis JR, Koch GG. The measurement of observer agree- ment for categorical data. Biometrics 1977;33:159-174.

15. Ferlay J, Soerjomataram I, Dikshit R, et al. Cancer incidence and mortality worldwide: sources, methods and major patterns in GLOBOCAN 2012. Int J Cancer 2015;136:E359-E386.

16. Jacobs MA, Wolff AC, Macura KJ, et al. Multiparametric and Multimodality Functional Radiological Imaging for Breast Cancer Diagnosis and Early Treatment Response Assessment. J Natl Cancer Inst Monogr 2015;2015:40- 17. Massat NJ, Dibden A, Parmar D, Cuzick J, Sasieni PD, 46.

Duffy SW. Impact of Screening on Breast Cancer Mortal- ity: The UK Program 20 Years On. Cancer Epidemiol Bio- markers Prev 2016;25:455-462.

18. Weedon-Fekjaer H, Romundstad PR, Vatten LJ. Modern mammography screening and breast cancer mortality: pop- ulation study. BMJ 2014;348:g3701.

19. Bahl M, Baker JA, Bhargavan-Chatfield M, Brandt EK, Ghate SV. Impact of Breast Density Notification Legisla- tion on Radiologists’ Practices of Reporting Breast Density:

A Multi-State Study. Radiology 2016;280:701-706.

20. Kim K, Song MK, Kim EK, Yoon JH. Clinical application of S-Detect to breast masses on ultrasonography: a study evaluating the diagnostic performance and agreement with a dedicated breast radiologist. Ultrasonography 2017;36:3- 21. Cho E, Kim EK, Song MK, Yoon JH. Application of Com-9.

puter-Aided Diagnosis on Breast Ultrasonography: Evalu- ation of Diagnostic Performances and Agreement of Ra- diologists According to Different Levels of Experience. J Ultrasound Med 2018;37:209-216.

22. Di Segni M, de Soccio V, Cantisani V, et al. Automated classification of focal breast lesions according to S-detect:

validation and role as a clinical and teaching tool. J Ultra- sound 2018;21:105-118.

Referințe

DOCUMENTE SIMILARE

Accuracy, specificity, sensitivity, positive predictive value (PPV) and negative predictive value (NPV) were calculated to show the diagnostic performance of S- Detect, the

In a study, Ma et al detected more blood flow and flow detail than color Doppler im- aging in evaluation malignant and benign breast lesions [17].. Similarly in our study, we

With the use of high-resolution ultrasound (US) equipment and the recent introduction of supplementary screening breast US in a population with dense breasts and an elevated risk

e) The doctoral thesis is officially submitted to the Doctoral School Secretariat, in printed and electronic format, together with the summary of the thesis in

The aim of this study was to evaluate the inter-observer agreement and effect of contrast injection on the visibility of liver lesions by radiologists reviewing ultrasound

Mammography has a low sensitivity and specificity in the diagnosis of papillary lesions since they may be ocult (especially benign papillomas) or may present with discrete

The evaluation of the CCLs in patients that underwent MG-guided stereotaxic biopsy revealed that the most common type of microcalcifications were amorphous-indistinct (52%, n=

It is clear that real-time intra-operative use of MRI and CT scans is not possible due to the inherent nature of these techniques. Myelography is one tool that can be used in