EXPERIMENTS AND SURFACES CHARACTERIZATION OF THE FEMORAL HEADS OF HIP PROSTHESES

GHEORGHE ION GHEORGHE, LILIANA-LAURA BADITA*

National Institute of Research and Development in Mechatronics and Measurement Technique, Bucharest, Romania

Total hip prostheses are necessary when the human hip joints are disturbed due to illness, under the negative influence of factors determined by the mode of human life or due to different accidents. Hip prostheses durability is influenced by the materials used to be made. Taking into account that is still complicated to increase hip prostheses durability, we started to characterize these prostheses in order to continue with nanomaterials coatings.

(Received October 20, 2011; Accepted February 28, 2012)

Keywords: Hip prostheses, Atomic force microscopy, Surface characterization, Tribological parameters

1. Introduction

Human body’s motor apparatus is composed of elements which allow mechanical interactions between human body parts and environment.

The joints are included in the interstitial elements system and are designed for transmitting fundamental components of the whole motor system’s movement. Joint is a system with passive components (such as articular ends of bones, cartilages, ligaments, synovial fluid) and active components (muscles). Kinematic parameters of the joint (e.g. the number of degrees of freedom) follow the shape of the bone surfaces in contact. The joint is a moving structure, but the ligaments are intended to limit mechanically the potential movement, to steer the surfaces in contact and to stiffen the joint. Depending on the shape of the contact surfaces that determine the potential of joint motion, there are:

- ball joints – which have spherical contact surfaces - cylindrical joints - contact area has cylindrical surface - ellipsoidal joints - contact surfaces are ellipsoidal - planar (sliding) - contact surfaces are parallel with a plan

- amphiarthrosis - contact surfaces are irregular and have minimum capacity of movement - combined - an anatomically autonomous joint is functional correlated with another joint and the movement in both occurs simultaneously.

The hip is essentially a ball and socket joint, formed by the articulation of the rounded head of the femur and the cup-like acetabulum of the pelvis. It forms the primary connection between the bones of the lower limb and the axial skeleton of the trunk and pelvis. The cup-like acetabulum forms at union of three pelvic bones and the joint may not be fully ossified under the age of 25 years. The large head of the femur attaches directly to the acetabulum. The head of the femur is attached to the shaft by a thin neck region that is often prone to fracture in the elderly, mainly due to the degenerative effects of osteoporosis.

Hip joint’s biomechanics. Due to the spherical shape, the joint has an inherent stability by its design. This does not alter the remarkable mobility of hip joint. In the joint, large forces are generated by the strong periarticular muscles, which balance the body’s weight amplified by the

*Corresponding author: [email protected]

(2)

bone levers. Any flaw in the joint mechanism changes the intra-articular forces distribution, causing degenerative changes.

Hip joint’s kinematics examines hip movements geometry apart from the force qualities it generates. Depending on the periarticular muscles, joint movements are performed in three planes of space and can be reduced to three pairs of fundamental movements performed around the three main axes of the joint. They intersect at the joint center of rotation, which corresponds to the center of the femoral head. These axes allow:

I. free extremity movements from the pelvis considered the fixed point;

II. movements of the pelvis to the free extremity fixed to the ground.

Considering the fixed point in the pelvis, the following possibilities can be distinguished for the movement of lower limb extremity from neutral-zero position (the human body anatomical position): flexion and extension movement (on or from the spine and on or from the thigh), abduction and adduction of the femur, internal (medial) and external (lateral) rotation of the pelvis, thigh or spine (Fig. 1).

Fig. 1. Movements in the hip joint.

Flexion and extension movement is performed to a transverse axis disposed in the frontal plane. The forward motion of the lower limb towards this axis achieves flexion (anteversion) and back movement achieves extension (retroversion). These two movements are the most important to the hip joint. Flexion’s amplitude is dependent upon the position of joint in the frontal plane (abduction and adduction). For example, the maximum flexion is achieved during easy abduction and neutral rotation positions. Passive movement is performed with highest amplitude and is limited by the contact between the soft parts. During normal movements of the hip joint are performed rarely pure movements after a single axis. Most times they are in the form of combinations of the three main types of movements.

Normal function of joints may be disturbed due to illness and under the negative influence of other factors determined by the mode of human life and professional activity. Osteoarthrosis is one of the most common forms of arthritis and is accompanied by gradual degradation of cartilaginous tissue. In such cases there is mutual friction of the bones, causing pain and accordingly result in partial or complete loss of joint mobility. Osteoarthritis is a disease in which the cartilages that lines and protects the joint surfaces deteriorate with the passage of time. This damage occurs as a result of chemical changes in the cartilage. Excessive body weight tensions joints, especially those that support weight, such as knees, hips and joints of the talus.

Osteonecrosis is a disease of bones, which occurs as a result of reducing or stopping blood irrigation of a certain portion of the bone. The disease causes the appearance of small cracks and defects in bones, which evolves over time and lead to bone destruction. Fractures of the femoral head may lead to un-consolidation and osteonecrosis, the appearance of decubitus caused by long horizontal position of the patient, ultimately causes death. Accidents produced by the simplest

(3)

ways c sensati pain, t consist A total

- - -

alloys acetabu torque decrea chance determ (determ mecha wear, t

the atta remov [3].

the su articul

hip pro

can have bla ions become

In such cas o restore lim ts in the rege l hip prosthe a cup that re a ball that w a stem that i

Joint comp or ceramic ular cup wit increased b ases the pres

es.

Joint elem mines the co

mined by hea The articul anical movem

The deteri third body w Abrasive w Adhesive w achment forc ed from one Third body Fatigue we urface. High

ating surface

2. New m So far it h ostheses: loo

asting results very pronou ses the drug mb length and

eneration of t esis (THP) co eplaces the hi will replace th s attached to

F

ponent of fe c (aluminum th much lar by the radiu ssure per un ment influenc

oefficient of ad’s diamete

lar surfaces m ments of the b

oration proc wear and fatig wear [2] repr

wear is prod ce is stronge surface and y wear refer ear can lead h subsurface

es leading to

materials f has not succe

osening and

s. In the fin unced.

treatment do d joints mob the joints fra onsists of thre

ip socket;

he fractured h o the shaft of

ig. 2. Compon

emoral prost m oxide or

rger diameter us difference nit area, red ces the wear f friction), b

r) and by the mainly deter body.

cess has a co gue wear [1]

resents the re duced where r than the yi is attached t rs to the inse

to subsurfac stresses ca accelerated

for hip pr eeded comple

fracture, reje

nal stages mo oes not give p ility is pseud agments dam ree parts (Fig head of the f f the bone to

nents of a tota

thesis is a p zirconium o er and thicke

e of compon ducing the c

r speed of th by the ang e type of lubr riorate due to omplex mech

(Fig. 1).

emoval of ma localized bo eld strength to the other.

ertion of a w ce cracks pr an also be fatigue wear

rostheses s etely removi ection physio

ost joint dise positive resu do-joints imp maged by dise

g. 2):

femur;

add stability

al hip prosthes

piece of met oxide). Pros er wall lead nents. Cup h chances of b he cup by th le of move rication that o the high su hanism, com aterial from o onding of the

of the mater wear particle opagating an caused by t r.

surfaces i ing the probl ological reac

eases cause ults, so the on

plantation. A ease using ps

y to the prosth

sis.

tal sphere m sthetic head

to the form having large bone resorpt he metal sur ment achiev allows it.

uperficial pre mbining abras one surface b e two surface rial. A small e between tw

nd flaking of third bodies

mprovem lems associa ctions of the

disability an nly way to a Arthroplasty s

seudo-implan

heses.

made from d d and polye mation of a e outdoor di

tion and loo rface finish ved in each essures produ sive wear, ad by the other.

es occurs, su piece of mat wo moving s ff of particle s between th

ment ated with the

body and th

nd pain lleviate surgery nts.

ifferent thylene friction iameter osening (which h cycle uced by

dhesive

uch that terial is surfaces es from he two

e use of he most

(4)

important, material wear. Therefore, it is necessary to have a resistant prosthesis, with anticorrosive composition and high mechanical properties.

Stainless steel, titanium alloy, polymers and ceramic composites undergo degradation after 10–15 years of use. Due to this reason, materials engineers must consider the physiologic loads placed on the implants. Material choices also must take into account immune system biocompatibility, the environment, corrosion issues, friction and wear of the articulating surfaces.

Sliding surfaces must be usually hard to keep the wear to minimum values. Ceramics, which have a very high strength, are not resistant to tearing and have a marked fragility. Metallic materials are relatively soft, but resistant to breakage. Thus, it is difficult to counteract adhesive and abrasive usages using the same material. Metallic biomaterials have different influences on the human body, is distinguishing different forms of biological reactions, according to: concentration of metal, exposure time and route of administration.

Generally, pure metals are rarely toxic. Toxic and allergenic effects depend on the concentration and nature mixtures (oxides, simple or complex salts). Two compounds of the same metal can induce strong responses, but different.

Cobalt-chrome alloy femoral heads articulating with UHMWPE acetabular liners have been the main stay of hip arthroplasty for more than 30 years. Zirconium alloy substrate is relatively soft when compared with cobalt-chrome alloy femoral heads and may deform in contact with acetabular shell materials in the case of dislocation. In the case of ceramics, with extremely low wear rates, remains a risk of catastrophic failure because of the inherently low fracture toughness of the materials. Oxinium components have a unique ceramic layer (4 µm thick) to mitigate the wear, akin to a zirconia ceramic material, and have an integral metal substrate that provides toughness and high fatigue strength.

Materials used for coatings.

International scientific community has made and still makes efforts to increase hip prostheses durability. In order to improve mechanical properties of hip prostheses, these have been coated with different materials, which have superior properties.

In order to improve mechanical properties of hip prostheses, these have been coated with different materials, which have superior properties.

Metal–matrix composite (MMC) coatings reinforced with hard ceramic particles are promising materials for improvement in various mechanical properties over conventional monolithic alloys.

A thin hydroxyapatite (HA) coating applied on textured metal surfaces [4], combines the mechanical attributes of metals with osteoconductive properties of metals and with the biocompatibility of ceramics [5].

Raw nanomaterials – which include nanoparticles and nanocrystalline materials that are readily manufactured and can substitute less performant bulk materials - can be also used as coatings in bone replacements, prostheses, and implants.

3. Experimental device

We studied the femoral heads surfaces of hip prostheses in order to see the level of wear.

Wear processes occurring inside different prostheses are an important source of debris, but these changes are often impossible to see with the naked eye. A methodology of ascending degrees of resolution was established using macroscopic, microscopic and nanoscale measurements. For these reasons we are trying to characterize femoral heads surfaces using a complex technique based on microscopic equipment [6].

We began with theoretical studies on sphere/plane thimbles, which simulated the femoral head and acetabular cup (Fig. 3). In the present case, the plane is the acetabular cup pressed by the spherical head, which destroys, in time, the prostheses [7]. Using this device and considering a vertical and static charge, we study the tribological properties of hip prostheses; determine the wear factor, the volume of material and the medium depth of the layer removed by wear process.

We realize hip prosthesis topography of these parts by using an atomic force microscope (NTEGRA Probe Microscope) working in the noncontact mode. Working principle of AFM is to

(5)

measur of a ca way w parame

Fig.

percen was id The la Signs observ occurr contac

Fig

areas w aspect degree head a be obs the rou movem

re the interac antilever with was obtained

eters.

. 3. Coated d

4. Resul Some resu The optica nt of the entir dentified an e arge scratch

of severe mi ve small scr red due to st ct interface.

g. 4. Images ob

Similar res were observe and areas w e, with a lot o and has an int

Using AFM served all the

ughness has ments of the b

ction force b h a pointed e the roughne

disks were teste

lts

ults obtained al microscopy

re surface ha embedded w could be rel icro-cutting w ratches and trong mechan

btained with A

sults were al ed on the fem with a small

of scratches.

termediate le M measurem e surface bum different val body that us

between tip a end. AFM im ess of the stu

ed articulating exper

d after our ex y study of a as isolated mi wear particle w

lated to the were detecte pits produc nical loading

AFM. Scratch corrosion fr

lso obtained moral head m

level of lum The third ca evel of lumin ments it can b

mps (Fig. 5, F lues (in diffe

es the prosth

and sample su mages were p

udied surface

a g against a ba rimental devic

xperiments a femoral head icro-scratche which scratc abrasion due ed also in an

ced by biot g combined

a

c hing (a), embed

retting (d) of fe

for the othe made from T minosity. In th ategory of re nosity. Here, be determin Fig. 6, Fig. 7 erent parts of heses.

urface using processed usi e and had be

all (polished s ce (b).

are presente d, made befo es despite of ched the fem e to third bo nother region

tribological with the loc

dded wear pa femoral head.

ers parts of t Ti. There are he second pa egions covers the deteriora ed exactly th 7). As it is sh

f the same fe

special mea ing Nova SP een calculate

b surface) on top

d in Fig. 4.

ore coating, r polished ma moral head su ody, namely (Fig. 4c). In corrosion. T cal temperatu

b

d rticle (b) micr

the femoral h e regions wit

art appears a s more than 5 ation level is he surface’s own in the fo emoral head)

asuring heads PM software.

ed other tribo

p of a disk (a)

revealed that acroscopic as urface (Fig. 4

the wear pa n Fig. 4d on The corrosio

ure increase

ro-cutting (c)

head. Three th a polished a high deteri 50% of the f s not so high roughness a following exa ), depending

s, made . In this ological

) and

t a high spect. It 4a, 4b).

articles.

e could on pits , at the

and

special d visual ioration femoral

. and can

amples, g on the

(6)

Fig. 5. Stainless steel surface microgeometry and tribological parameters measured using NTEGRA Atomic Force Microscopy

Fig. 6. Ti6Al4V surface microgeometry and tribological parameters measured using NTEGRA Atomic Force Microscopy

(7)

Fig. 7. CoCr surface microgeometry and tribological parameters measured using NTEGRA Atomic Force Microscopy.

After few experimental results it was observed that the friction is highest in tests with stainless steel and slightly lower for Ti6Al4V. Taking into account these measurements, the femoral heads surfaces will be coated in order to observe the favourable coating material.

5. Conclusions

As it was demonstrated in the last years, femoral head damage may occur during hip joint movements and may lead to accelerated materials wear. In order to obtain a clear characterization of the femoral head surfaces the study of its topography is useful. Such a study can be made by different techniques, but we decided on AFM because its images display high quality and dense nanocrystalline structure of prepared thin films.

Taking into account the obtained results we shall continue the research and AFM microscopy studies. The next steps will be based on femoral heads surfaces coatings that offer the opportunity to improve system’s properties. Different materials will be used and characterized to protect against hip prosthesis breakdown.

References

[1] G. Bergmann, F. Graichen, A. Rohlmann, N. Verdonschot, G. H. van Lenthe, Journal of Biomechanics, 34, 421 (2001)

[2] Z. L. Dong, K. A. Khor, C. H. Quek, T. J. White, P. Cheang, Biomaterials, 24, 97 (2003) [3] Gh. I. Gheorghe, L. L. Badita, Advanced micro and nanotechnologies in mechatronics, Ed.

CEFIN, Bucharest, 2009 (in Romanian).

[4] Y. Liu, X. Dan, Y. Du, F. Liu, Journal of Materials Science, 39(12), 4031 (2004)

[5] C. M. Lin, S. K. Yen, Journal of Materials Science: Materials in Medicine, 16(10), 889 (2005) [6] Gh. I. Gheorghe, Integrated intelligent MicroNano- Technologies, Ed. CEFIN, Bucharest, 2010, (in Romanian)

[7] L. Capitanu, The Orthopaedical Endoprostheses durability, BREN, Bucharest, 2004 (in Romanian)

Referințe

DOCUMENTE SIMILARE

In this research an investigation was made on the mechanical, thermal, electrical properties of E-glass fiber reinforced epoxy composites dispersed with fine iron (III)

In order to improve the response characteristics of the device, 3-aminopropyltrimethoxysilane (APTMS) was deposited onto the IZO film to improve its surface state,

A polymer matrix protects the optical fiber and preserves its mechanical properties, while the magnetic particles in composite have functional role - sensing of external

Additionally, related properties, i.e., membrane forming ability, water resistance, ion stability, mechanical stability and storage ability of nanocomposite emulsion in the

We prepared POC-ZnO nano-composite scaffolds with variation of ZnO concentration and a series of experiments was conducted in order to: (i) investigate the influence of ZnO

The aim of this research was to evaluate the effect of chitosan and microfluidization pretreatment on mechanical and microstructural properties of mung bean

Further the influence of exposure to sonication on PS-GO composite materials morphology, structure, mechanical and thermal properties was investigated... Materials

In this article, molecular dynamics simulation is used to study the mechanical properties of graphene sheets under uniaxial tensile loading..