Review

Diagnostic performance of ultrasound for rotator cuff tears:

a systematic review and meta-analysis

Weiyu Liang

*

, Huaiyu Wu

*

, Fajin Dong, Hongtian Tian, Jinfeng Xu

* the authors shared the first authorship

Department of Ultrasound, First Affiliated Hospital of Southern University of Science and Technology, Second Clini- cal College of Jinan University, Shenzhen Medical Ultrasound Engineering Center, Shenzhen People’s Hospital, Shenzhen, China

Received 03.11.2019 Accepted 02.02.2020 Med Ultrason

2020, Vol. 22, No 2, 197-202 Corresponding author: Jinfeng Xu

Department of Ultrasound, First Affiliated Hospital of Southern University of Science and Technology, Second Clinical College of Jinan University, Shenzhen Medical Ultrasound Engineering Center, Shenzhen Public Service Platform on Tumor Precision Medicine and Molecular Diagnosis

E-mail: [email protected]

Introduction

In the human body the shoulder can be seen as the most flexible and vulnerable joint. The primary causes of pain in shoulder and dysfunction are rotator cuff (RC)

avulsions [1,2]. More than 72% of all shoulder pain can be attributed to RC pathology [3,4]. Due to the difficulty of confirming the cause of pain in the shoulder by clini- cal examination [5], medical imaging tends to be used to verify the diagnosis [1,6]. It is hard to achieve differ- ential shoulder pathologies’ diagnosis because many im- pairments, such as instability and shoulder impingement, could have similar symptoms [7,8]. Therefore, an accu- rate differential diagnosis is important for clinical deci- sion making [9]. Particularly, the accurate RC tears’ size and location measurement is important for the efficiency of pre-operative planning [10].

Various clinical examinations, such as arthrography, magnetic resonance imaging (MRI) and ultrasound, have been used to evaluate RC [11]. Both arthrography and MRI are the reference standards for diagnosing RC tears Abstract

Aim: Rotator cuff (RC) tears are considered to be the main reason for shoulder pain. Although ultrasound is a useful method to detect it, its effectiveness when diagnosing RC tears has been a heated discussion. Therefore, we aimed at evaluat- ing RC tears’ ultrasound accuracy by conducting a systemically review and pooled comprehensive analysis. Materials and methods: Relevant articles up to May 2018 were searched from the Cochrane Library, Embase, and Pubmed databases. Either arthroscopy or magnetic resonance imaging (MRI) was considered as a reference standard. The results were estimated by pooled-sensitivity (P-SEN), pooled-specificity (P-SPE), pooled-diagnostic odds ratio (P-DOR), pooled-likelihood ratio+ (P- LR+), pooled-likelihood ratio- (P-LR-), and the area under the summary receiver operating characteristic (SROC). Result: We selected seven prospective studies in accordance with the inclusion criteria that covered 554 rotator cuff tears in 868 patients.

The P-SEN, P-SPE, P-LR+, P-LR-, P-DOR, area under the SROC curve of diagnostic performance of ultrasound for RC and post-test probability were 95% (95% CI: 88 – 98), 72% (95% CI: 61 – 81), 3.41 (95% CI: 2.40 – 4.84), 0.08 (95% CI: 0.03 – 0.16), 45.31 (95% CI: 21.04 – 97.59), 89% (95% CI: 0.86 – 0.91), and 46% and 2%, respectively. Conclusion: Our meta- analysis demonstrates that ultrasound has a high efficiency for RC tears’ diagnosis. It can be a promising method in patients with suspected RC tears because of its high sensitivity, specificity and diagnostic accuracy.

Keywords: ultrasound; diagnosis; meta-analysis; systematic review; rotator cuff tears.

DOI: 10.11152/mu-2352

(2)

[12]. However, arthroscopy is an invasive method of in- spection that is not widely accepted and MRI is time- consuming and expensive [1,9]. Many studies have indi- cated that ultrasound is a useful imaging method to detect RC tears [13-18]. Compared to arthroscopy and MRI, ul- trasound is relatively cheap, noninvasive, and permits a dynamic assessment [19].

Examination of the RC structure by ultrasound was first reported in the late 1970’s [20]. With the emergence of high-frequency probes, the application of ultrasonog- raphy in the examination of human fine structure has become more extensive. There are many published stud- ies on the use of ultrasound to diagnose RC tears, which have been divided into either complete or partial tears, and the diagnostic efficiency of ultrasound may be differ- ent in each [21]. There has been one systematic review assessing the sensitivity of ultrasound in detecting shoul- der injuries [9]: however, this study separately evaluated these two types of tears. At present, no research has been undertaken on both complete and partial RC tears. There- fore, the purpose of this meta-analysis was to evaluate both types of tears in order to decide the diagnostic po- tential of ultrasound for RC tears.

Materials and methods Search strategy

All relevant studies from the Cochrane Library, Em- base, and Pubmed databases were retrieved. Literature was retrieved using MeSH and free words and English publications limited it. The search strategies applied by the Cochrane Library, Embase, and Pubmed are as fol- lows: (infraspinatus AND tear) or (supraspinatus AND tear) or (tears AND minor AND tear) or (subscapularis AND tear) or (shoulder AND joint AND abnormalities) and ultrasound.

Study selection

We included all studies that used ultrasound to evalu- ate rotator cuff tears, had arthroscopy or MRI as the diagnostic criterion and directly or indirectly provided primordial data that could be used to calculate negative likelihood ratios (LR−), positive likelihood ratio (LR+), specificity and sensitivity. Cadaveric and animal trials, single case reports, incomplete research, meta-analyses and systematic reviews, studies using a population of fewer than 20 patients and research which did not have a comparison of reference standard were excluded. Two authors with the same level of expertise and experience selected literature meeting the inclusion criteria after reading the full text. A third author reviewed the search results and the assessment process and solved any dis- cordances.

Data extraction and quality assessment

After collecting all qualified studies, we extracted the following content: reference standard, number of rotator cuff tears for analysis, number of patients, study design, year of publication, the surname of the first author, TP, FN, FP, TN, sensitivity and specificity.

The QUADAS-2 (Quality Assessment of Diagnostic Accuracy Studies-2) tool [22] was used to evaluate the quality of the eligible studies. It was made up of four parts: patient selection, index test, reference standard, and flow and timing. Every part was evaluated for bias risk and assigned a score of unclear, high, low. The refer- ence standard domains, patient selection, index test were measured for the presence of attention on the applica- bility, and were likewise given a score of low, high, or unclear. If a research is evaluated by “low” in all parts regarding applicability or bias, it is good to comprehen- sively judge the “low risk of bias” or “low concern re- garding applicability” for the research. If a research study is evaluated by “high” or “unclear” in one or more parts, it can be judged “at risk of bias” or “concerns regarding applicability”.

Statistical analysis

We extracted the data from the included literature and recorded it using a four-fold table. We analyzed the data using RevMan 5.3 and Stata 12.0. RevMan 5.3 was applied in order to measure the eligible studies’ meth- odological quality. It used stata to draw statistical graphs and pool statistical indexes such as forest plots for sensi- tivity, specificity, diagnostic odds ratio (DOR) with cor- responding 95% confidence intervals (CI), and the area under summary receiver operating characteristic curve (SROC). Heterogeneity of the included references was verified after statistical analysis. We used the inconsist- ency index (I2) to estimate the heterogeneity across the included studies. If the I2 was <0.5, in the included stud- ies, no heterogeneity was found. On the contrary, it in- dicated the heterogeneity’s result. If heterogeneity could not be omitted, we applied a random effects model, which evaluates the combined effect’s influence containing all information (sensitivity, specificity, accuracy, positive predictive value, negative predictive value, etc) [23].

The place beneath the ROC’s curve shows the diagnostic test’s accuracy, along with the lower accurate degree de- fined as an area of 0.5 - 0.7, certain accuracy as an area of 0.7 - 0.9, and higher accuracy with a place >0.9 [24].

Publication bias

The Deeks’ funnel plot asymmetry test was used to evaluate potential publication bias. It was applied by regressing diagnostic log odds ratio (lnDOR) versus in- verse of the square root (1/sqrt) of the effective sample size (1/ESS1/2) and measured through effective sample

(3)

size. A p-value <0.10 for slope coefficients demonstrated the key asymmetry.

Results

Figure 1 shows the study selection process’s flow chart. On the basis of the previously determined strat- egy of search, we searched an overall of 2,356 articles.

Among these, 2,040 were from Pubmed, 273 were from Embase, and 43 were from the Cochrane Library. There were 145 duplicate materials that were excluded using ENDNOTE X7. There were 2,198 studies excluded ac- cording to the title and abstract, including eight reviews systematically and/or meta-analysis and 166 case reviews or/and researches. For the remaining 13 articles, we ex- cluded six of them after reading the full text because they could not provide four-fold data directly or indirectly. In the end, the meta-analysis included seven articles. The included studies’ features are listed in Table I.

According to the methodological assessment from the QUADAS-2 checklist written by RevMan 5.3, patient selection, index test, reference standard, and flow and timing indicated good quality. The results are detailed in figure 2. The selected studies’ quality was relatively high and four of them had QUADAS full scores [11,12,21,25].

Table I. Characteristics and diagnostic performance of included studies Author (year) Country Number of

patients (men)

Full thickness (partial thickness) rotator cuff tears

Reference TP FP FN TN SE (%) SP

(%)

Abd-ElGawad (2013) [11] Egypt 40 (27) 23 (13) Arthroscopy 35 2 1 2 97 50

Cole (2016) [12] Australia 238 (unclear) 139 (77) Arthroscopy 197 3 19 19 91 86

Elmorsy (2017) [19] UK 125 (58) 48 (13) Arthroscopy 47 11 14 53 77 83

Co (2012) [21] Canada 37 (22) 12 (10) MRI 20 6 2 11 91 65

Singisetti (2011) [25] UK 82 (56) 55 (6) Arthroscopy 55 19 0 22 1 54

Jeyam (2008) [26] UK 64 (unclear) 16 (6) Arthroscopy 21 1 1 6 95 86

Ziegler (2004) [27] USA 282 (173) 261 (1) Arthroscopy 261 1 14 6 95 86

SE: Sensitivity, SP: Specificity, TP: True positive rate, FP: False positive rate, FN: False negative rate, TN: True negative rate

Fig 2. Quality assessment using the Quality Assessment of Diagnostic Accuracy Studies-2 (QUADAS-2) tool: A) Risk of bias and applicability concerns graph: authors’ ratings for each domain are presented as percentages across included studies;

B) Risk of bias and applicability concerns summary: authors’

ratings for each domain for each included study.

Fig 1. Flowchart of the study selection process

(4)

There are three articles in which the author did not men- tion the time interval during the ultrasound examination and the reference standard’s implication [19,26,27].

The random effects model was used to combine re- search data of eligible studies because the sensitivity and the LR− with 95% CI and forest plots (fig 3a,b) displayed obvious heterogeneity, together with an I2 of 79.73% and

73.80%. Figures 3, 4a,b shows that the pooled sensitivity, specificity, LR+, LR-, DOR, area under the SROC curve of diagnostic performance of ultrasound for rotator cuff, and post-test probability were 95% (95% CI: 88 – 98), 72% (95% CI: 61 – 81), 3.41 (95% CI: 2.40 – 4.84), 0.08 (95% CI: 0.03 – 0.16), 45.31 (95% CI: 21.04 – 97.59), 89% (95% CI: 0.86 – 0.91), and 46% and 2% respec-

Fig 4. a) Fagan’s Nomogram with pretest probability at 20%; b) Hierarchical summary receiver operating characteristic curve (HSROC) with summary point, summary estimates, 95% confidence region and 95% prediction region for all included studies of ultrasound for evaluating rotator cuff tears (n =7); c) Deeks’ funnel plot with superimposed regression line for identifying publica- tion bias, Log Odds Ratio versus 1/sqrt (effective sample size) (Deeks) indicates that no significant bias was found. ESS = effective sample size.

Fig 3. Forest plot for a) sensitivity and specificity of ultrasound for evaluating rotator cuff tears; b) a positive and negative likelihood ratio of ultrasound for evaluating rotator cuff tears; c) diagnostic odds ratio of ultrasound for evaluating rotator cuff tears.

(5)

tively. Figure 4c is a Deeks’ funnel plot, which was used to indicate a systematic error in publication bias. The p- value was 0.68 and regression line’s both sides distrib- uted the included studies, indicating that the publication bias was not apparent.

Discussion

The rotator cuff is an important structure that main- tains the stability of the shoulder joint and is important in the abduction, external rotation and shoulder’s internal rotation. It is the only tendonous tissue between the two bones of the shoulder joint, and is easily damaged due to its particular anatomical position and repetitive move- ment [28]. Many methods can be used to examine the shoulder joint. Radiography and CT are widely used in the diagnosis of bone damage and its extent, but the di- agnosis of soft tissue injury is inferior to ultrasound and MRI. Although MRI is currently the most reliable, non- invasive diagnostic method for rotator cuff tears evalu- ation, it is an expensive imaging technique, has low re- producibility and is not suitable for use in patients with metal-mounted devices [12].

In the recent years, with the development of ultra- sound technology and the emergence of high-frequency probes, ultrasound was increasingly used for musculo- skeletal systems pathology. Many studies have shown that there are not obvious differences between ultrasound and MRI when detecting rotator cuff tears [19,26,29].

However, though ultrasound is inexpensive, noninvasive and convenient, its effectiveness when diagnosing RC tears is still controversial. We found that the combined sensitivity, specificity, LR- and DOR of the included studies show higher heterogeneity (I2 >50%). Even so, ultrasound had a high sensitivity and credible specific- ity when diagnosing the rotator cuff tear. We also found a high LR+, DOR, certain accuracy and a lower LR−.

The LR is a composite indicator reflecting sensitivity and specificity, which can better reflect the diagnostic value of diagnostic experiments [30].

Although we systematically evaluated the inclusion of literature using a rigorous process, there are still de- ficiencies. For example, Elmorsy et al did not mention the time interval between the ultrasound exam and the reference standard’s implementation [19], which could generate the bias of patient flow. In addition, our study did not contain unpublished and non-English articles. Fi- nally, the studies we included are from different regions, which is likely to produce differences in compliance. In all, compliant bias, language, and publication are shown.

This meta-analysis demonstrates that combined sen- sitivity and LR- are of a high degree of heterogeneity and

we have three explanations for this. First, a portion of the included studies was selected for patients with clinically, highly suspected rotator cuff tears [19,21]. Second, some papers included in our study individually analyzed partial and complete tears [25-27], but we combined the two for analysis. Third, the frequency of the probe used and the diagnostic level of the operating physician was different in each study. These factors are likely to introduce bias and represent limitations to this study.

Conclusion

In conclusion, our meta-analysis demonstrates that ultrasound has a high efficiency of diagnosis for rotator cuff tears. As for patients, it can be a promising approach with suspected rotator cuff tears due to its high diagnostic accuracy and sensitivity specificity.

Conflicts of interest: none

Acknowledgements: The project was supported by the National Natural Science Foundation of China (NO.81771841) and the Commission of Scientific and Technology of Shenzhen (No. JCYJ20170307095706 970).

References

1. Milosavljevic J, Elvin A, Rahme H. Ultrasonography of the rotator cuff:a comparison with arthroscopy in one-hundred- and-ninety consecutive cases. Acta Radiol 2005;46:858- 2. Seibold CJ, Mallisee TA, Erickson SJ, Boynton MD, 865.

Raasch WG, Timins ME. Rotator cuff: evaluation with US and MR Imaging. Radiographics1999;19:685-705.

3. Macfarlane GJ, Hunt IM, Silman AJ. Predictors of chron- ic shoulder pain:a population based prospective study. J Rheumatol 1998;25:1612-1615.

4. Mitchell C, Adebajo A, Hay E, Carr A. Shoulder pain:

diagnosis and management in primary care. BMJ 2005;331:1124-1128.

5. Naqvi GA, Jadaan M, Harrington P. Accuracy of ultra- sonography and magnetic resonance imaging for detec- tion of full thickness rotator cuff tears. Int J Shoulder Surg 2009;3:94-97.

6. Roberts CS, Walker JA 2nd, Seligson D. Diagnostic ca- pabilities of shoulder ultrasonography in the detection of complete and partial rotator cuff tears. Am J Orthop (Belle Mead NJ) 2001;30:159-162.

7. Mack LA, Gannon MK, Kilcoyne RF, Matsen RA 3rd.

Sonographic evaluation of the rotator cuff. Accuracy in patients without prior surgery. Clin Orthop Relat Res 1988;234:21-27.

8. Friedman RL, Hidalgo HJ, Gilmer PW, et al. Ultrasonogra- phy of the rotator cuff: analysis of results in a community setting. J Shoulder Elbow Surg1993;2:22-6.

(6)

9. Smith TO, Back T, Toms AP,. Hing CB. Diagnostic accura- cy of ultrasound for rotator cuff tears in adults: A systematic review and meta-analysis. J Clin Radiology 2011;66:1036- 1048.

10. Farin PU, Kaukanen E, Jaroma H, Väätäinen U, Miettinen H, Soimakallio S. Site and size of rotator-cuff tear. Findings at ultrasound, double-contrast arthrography, and computed tomography arthrography with surgical correlation. J Invest Radiol 1996;31:387-394.

11. Abd-ElGawad EA, Ibraheem MA, Fouly EH. Evaluation of supraspinatus muscle tears by ultrasonography and magnet- ic resonance imaging in comparison with surgical findings.

Egypt J Radiol Nucl Med 2013;44:829–834.

12. Cole B, Twibill K, Lam P, Hackett L, Murrell GA. Not all ultrasounds are created equal: general sonography versus musculoskeletal sonography in the detection of rotator cuff tears. Shoulder Elbow 2016;8:250–257.

13. Allen GM, Wilson DJ. Ultrasound of the shoulder. Eur J Ultrasound 2001;14:3-9.

14. Dinnes J, Loveman E, McIntyre L, Waugh N. The effective- ness of diagnostic tests for the assessment of shoulder pain due to soft tissue disorders: a systematic review. Health Technol Assess 2003;7:1-166.

15. Jacobson JA, van Holsbeeck MT. Musculoskeletal ultra- sonography. Orthop Clin North Am 1998;29:135-167.

16. Middleton WD, Teefey SA, Yamaguchi K. Sonography of the rotator cuff: analysis of interobserver variability. AJR Am J Roentgenol 2004;183:1465-1468.

17. Martinoli C, Bianchi S, Prato N, et al. US of the shoulder:

non-rotator cuff disorders. Radiographics 2003;23:381-401.

18. de Jesus JO, Parker L, Frangos AJ, Nazarian LN. Accuracy of MRI, MR arthrography, and ultrasound in the diagnosis of rotator cuff tears: a meta-analysis. AJR Am J Roentgenol 2009;192:1701-1707.

19. Elmorsy A, Keightley A, Flannery M. Accuracy of Ultra- sonography (US) and Magnetic Resonance Imaging (MRI) in Detection of Rotator Cuff Tears in District General Hos- pital. Pol J Radiol 2017;82: 634-637.

20. Seltzer SE, Finberg HJ, Weissman BN, Kido DK, Collier BD. Arthrosonography: gray-scale ultrasound evaluation of the shoulder. Radiology 1979;132:467-468.

21. Co S, MD, Bhalla S, Rowan K, Aippersbach S, Bicknell S.

Comparison of 2- and 3-dimensional shoulder ultrasound to magnetic resonance imaging in a community hospital for the detection of supraspinatus rotator cuff tears with improved worktime room efficiency. Can Assoc Radiol J 2012:63:170-176.

22. Whiting PF, Rutjes AW, Westwood ME, et al. QUADAS-2:

A revised tool for the quality assessment of diagnostic ac- curacy studies. Ann Intern Med 2011;155:529-536.

23. Higgins JP, Thompson SG, Deeks JJ, Altman DG. Meas- uring inconsistency in meta-analyses. BMJ 2003;327:557- 24. Swets JA. Measuring the accuracy of diagnostic systems. 560.

Science 1988;240:1285–1293.

25. Jeyam M, Funk L, Harris J. Are shoulder surgeons any good at diagnosing rotator cuff tears using ultrasound?: A comparative analysis of surgeon vs radiologist. Int J Shoul- der Surg 2008;2:4–6.

26. Singisetti K, Hinsche A. Shoulder ultrasonography versus arthroscopy for the detection of rotator cuff tears: analysis of errors. J Orthop Surg (Hong Kong) 2011;19:76-79.

27. Ziegler DW. The use of in-office, orthopaedist-performed ultrasound of the shoulder to evaluate and manage rota- tor cuff disorders. J Shoulder Elbow Surg 2004;13:291- 28. Moosmayer S, Heir S, Aaser P, Smith HJ. Ultrasound ex-297.

amination of the shoulder a method description.Tidsskr Nor Laegeforen 2004;124:177-180.

29. Modi CS, Smith CD, Ho K, et al. Accuracy of high- resolution ultrasonography in the diagnosis of articular side partial thickness rotator cuff tears. Shoulder Elbow 2010;2:267–270.

30. Bürkner PC, Doebler P. Testing for publication bias in diagnostic meta-analysis: a simulation study. Stat Med 2014;18:3061-3077.

Referințe

DOCUMENTE SIMILARE

MT – medial tubercle of the pos- terior process of the talus; LT – lateral tubercle of the posterior process of the talus; star - groove between the two tubercles of the

Aims: In the present study, a meta-analysis was performed to evaluate the diagnostic value of endobronchial ultrasound (EBUS) elastography for differentiating benign and

Faced with the possible insurrection of the body and of the sensible in general, and in order to surpass the possible nefarious consequences of such a logical dead end, (clear

The following sections present the syntax for defining fuzzy variables, using fuzzy variables in LHS patterns and in facts, declaring certainty factors, changes made to the

This sensitivity analysis showed that even when only pathologically confirmed cases were included, the diagnostic performance for distinguishing HP from other SETs

Adnexal masses dif- ficult to classify as benign or malignant using subjective assessment of gray-scale and Doppler ultrasound find- ings: logistic regression models

We calculated the summary statistics for sensitivity (Sen), specificity (Spe), positive and negative likelihood ratio (LR + /LR − ), diagnostic odds ratio (DOR) and receiver operating

All articles identified as potentially eligible by either reviewer were reviewed as abstracts or full-text articles. Eligible studies met the following inclusion criteria: 1)