Original paper

DOI: 10.11152/mu.2013.2066.183.sko

Abstract

Ultrasound examinations are recognised as being safe. The greatest epidemiological threat during the performance of examination is the transfer of pathogenic and opportunistic microorganisms between patients and from personnel to patients.

Colonization of the skin with opportunistic bacteria of immunocompromised and high risk patients may lead to infection fol- lowing an ultrasound scan. Aim: To identify and evaluate the strains of bacteria occurring on ultrasound equipment subjected to unexpected control performed by a local infectious diseases control team. We assumed that transducers, gel holders and gel bottles can be contaminated with normal human skin microflora. The remaining tested parts of the ultrasound equipment could possibly be contaminated with normal human skin microflora and other pathogens. Material and methods: The swabs were taken from ultrasound scanners located in various hospital settings, from out-patient based radiology scanning rooms to operating theatre, and cultured. Results: Among all isolated 23% strains were classified as environmental microflora; 8% as strains related to patient’s skin contamination; and 13 % strains constituted pathogenic Gram-negative rods.. The remaining strains were classified as opportunistic flora 38%. High prevalence of opportunistic bacteria cultured in our study lead to the modification of the ultrasound cleaning procedures in both institutions and recommendation of the use of antibacterial wipes to clean all parts of ultrasound equipment in contact with patients’ skin and examiners. Conclusions: Contamination not only affects parts of diagnostic equipment placed in direct contact with the patient, but also, those surfaces that only medical per- sonnel have had contact with.

Keywords: ultrasonography, cross infection, bacterial contamination

Can diagnostic ultrasound scanners be a potential vector of opportunistic bacterial infection?

Paweł Skowronek

1

, Artur Wojciechowski

2,3

, Piotr Leszczyński

4

, Paweł Olszewski

3

, Marcin Sibiński

5

, Michał Polguj

6

, Marek Synder

5

1Orthopaedics and Traumatology Department, The Jan Kochanowski University, Kielce, Poland, 2St. George’s University Hospital, London, UK, 3Department of Radiology, Medical University of Warsaw, Poland, 4Nosocomial Infection Control Team, Warsaw Medical University, The Infant Jesus Teaching Hospital, Warsaw, Poland, 5Clinic of Orthopedics and Pediatric Orthopedics, Medical University of Lodz, Poland, 6Department of Angiology, Chair of Anatomy. Medical University of Lodz, Poland

Received 04.01.2016 Accepted 20.03.2016 Med Ultrason

2016, Vol. 18, No 3, 326-331

Corresponding author: Artur Wojciechowski

St. George’s University Hospital, Cranmer Terrace, London SW17 0RE London, UK

E-mail: [email protected]

Introduction

A variety of imaging techniques are used in the di- agnostic process or for monitoring the treatment pro- cess, and their use is becoming more common, with significant annual increases of the number of imaging examinations performed worldwide [1]. Personnel and patients are aware of the risks associated with exposure

to the ionising radiation at the time of X-ray, computed tomography, vascular studies as well as possible interfer- ence of the magnetic field with heart pacing devices and metallic body implants in patients subjected to magnetic resonance imaging. However, ultrasound examination is regarded as non-invasive, and the patients assume no additional risk while undergoing it is involved [2]. The greatest epidemiological threat during the performance of the ultrasound examination is the transfer of pathogen- ic microorganisms between patients and from personnel to patients that may lead to infection with alert pathogens as well as infection with opportunistic bacteria.

The aim of this study was to identify and evaluate the strains of bacteria occurring on ultrasound equip- ment subjected to decontamination procedures recom-

(2)

mended by manufacturers. We developed the hypothesis that contamination not only affects parts of the diagnostic equipment placed in direct contact with the patient, but also those surfaces that only medical personnel have had contact with.

Materials and methods

The prospective study was performed in two hospi- tals in June 2014: The Jan Kochanowski University Hos- pital in Kielce referred to as centre A and 1st Department of Clinical Radiology Medical University of Warsaw re- ferred to as Centre B. The following ultrasound scanners were submitted to control: in Centre A – Esaote Mylab Class C and Esaote Mylab Seven at main ultrasound room and Esaote Mylab Twice at surgical department operating theatre; in Centre B – two Logiq 9 GE at main ultrasound room, ATL3000 Philips at transplantology de- partment ultrasound room, Logiq 5 GE at intensive care unit and Xario Toshiba at surgical department operating theatre. In both hospitals the ultrasound machines in out- patient settings – in radiology department, transplantol- ogy department were sampled at the end of ultrasound session. The machines in operating theatre and at inten- sive care unit were sampled before use. Sterile swabs that have been moistened with sterile water were used for ma- terial collection. Samples were taken from the following elements of ultrasound scanners: convex transducer – the scanhead scanning surface and 60 – 80 cm connecting cable length proximal to from the transducer, keyboard, touch screen/controls, gel containing bottle, and gel bot- tle holder. The scanners were located at Radiology De- partment, the operating theatre of the General Surgery Department, the transplantation clinic and the intensive care department in two Hospitals.

Swabs were taken in both centres. A total number of 39 swabs were taken: site 1 – convex transducer – scan- ning surface and cable – 8 swabs, site 2 – keyboard – 8 swabs, site 3 – touch screen and controls – 8 swabs, site 4 – gel bottle – outer surface and area around opening – 7 swabs, site 5 – gel bottle holder – 8 swabs. One sampling site was omitted in Centre B, that is the ultrasound scan- ner in the operating theatre, because sterile, disposable ultrasound gel was used there, not gel stored in the bot-

tles. The swabs were put into the transfer medium and sent to the microbiology laboratory within the hour. All samples were taken by a competent sampling officer who had been trained to carry out this procedure.

Swab alone is not highly sensitive for detecting the bacteria present on tested surfaces. To increase the sensi- tivity of the test, the samples were seeded on a Schaedler liquid base [3] incubated under laboratory conditions for 48 hours, and after growth was observed, they were in- oculated seeded on blood agar, McConkey medium and chromogenic media to isolate MRSA, ESBL and VRE strains (bioMeriux, Poland). Strains were identified on the basis of cell and colony morphology [4] and addi- tional tests. In addition, tests intended to identify epide- miological markers were carried out.

Results

A total of 39 swabs were taken from 8 ultrasound machines in two centres in June 2014. Fifteen swabs were taken in centre A and 24 swabs in centre B. All swabs were cultured. In centre A Gram-positive bacteria grew on 15 swabs and Gram-negative bacteria grew on 2 swabs. In centre B 9 swabs showed growth of Gram- positive bacteria and 3 Gram-negative (table I). No growth was observed in 12 swabs taken in centre B. In both centres Gram-negative bacteria were isolated only from swabs collected form equipment belonging to Ra- diology Department 2 in centre A and 3 in centre B. The isolated rods belonged to the Pantoea and Pseudomonas (P. stutzeri) or Achromobacter spp. The cultured strains were not ESBL (Extended spectrum beta lactamase).

No carbapenem non-susceptibility was observed. While Meticillin-Resistant Coagulase-Negative Staphylococci belonging to genera: Staphylococcus haemolyticus and S.

hominis were isolated from 11/15 swabs in Centre A, no such staphylococci were isolated in Centre B. Meticillin- susceptible S. aureus was isolated only in Centre A. No isolates of meticillin-resistant S. aureus were isolated in either centre. Similarly, no glycopeptide-resistant entero- cocci were isolated.

Spores forming rods Bacillus spp. were isolated in both centres with roughly equal frequencies: 3/15 swabs in A and 7/24 swabs in B (table II).

Table I. Comparison of positive swab cultures of ultrasound scanners in radiology departments and in other departments.

Centre Bacteria In Radiology Department In other Departments

A Gram-positive 10/10 5/5

Gram-negative 2/10 0/5

B Gram-positive 7/10 2/14

Gram-negative 3/10 0/14

The results are expressed as the number of positive swab culture/total number of samples

(3)

Table II. Strains of isolated bacteria in both Centres Sampling siteBacteriaCentre A Radiology Dept. scanner

No 1

Centre A

Radiology Dept. scanner No 2

Centre A

Surgery Dept. Operating theatre scanner

Centre B

Radiology Dept. scanner No 1

Centre B

Radiology Dept. scanner No 2

Centre B

Transplantology Dept. scanner

Centre B

Surgery Dept. Operating theatre scanner

Centre B ICU scanner 1. Convex trans- ducer – scan-

ning surface and cable Gram positiveS. haemolyti- cus MRS. hominis MR Bacillus spp..- three pheno

- types

Bacillus spp.MSCNSNo growthNo growthNo growth Gram negative No growthNo growthNo growthNo growthNo growthNo growthNo growthNo growth 2. Keyboard Gram positive

S. aureus MS S. haemolyti

-

cus MR Corynebacte

- rium spp.

S. haemolyti-

cus MR S. epidermidis MS

S. hominis MREnterococcus spp.Bacillus spp.No growth No growth No growth Gram negativeNo growthNo growthNo growth rods ESBL (-) rods ESBL (-)No growthNo growthNo growth 3. Touch screen and controlsGram positiveS. haemolyti- cus MR E. faecium

S. haemolyti- cus MR E. hirae

S. aureus MSNo growth No growth

MSCNS, Bacillus spp.

No growth No growth Gram negative Rods ESBL (-)Rods ESBL (-)No growthNo growthNo growthNo growthNo growthNo growth 4. Gel bottle – outer walls, area around the mouth

Gram positiveS. hominis MRCorynebacte- rium striatumS. hominis MR E. faecalis, Corynebacte- rium spp.

No growth

MSCNS, Bacillus spp.

No growth NCNo growth Gram negativeNo growthNo growthNo growthRods ESBL(-)No growthNo growthNCNo growth 5. Gel bottle holder Gram positiveS. hominis MRKocuria kris- tinaeS. haemolyti-

cus MR S. hominis MR

Bacillus spp.Bacillus spp.No growth No growth Bacillus spp. Gram negativeNo growthNo growthNo growthNo growthNo growthNo growth No growthNo growth MR meticillin resistant, MS meticillin susceptible, MRCNS Meticillin-Resistant Coagulase-Negative Staphylococcus, ESBL: Extended Spectrum Beta Lactamase, NC not cul- tured (no swab was taken, disposable sterile US gel is used in operating theatre settings in site B)

(4)

Isolated bacteria were classified as environmental flora – Bacillus spp. rods – isolated from 9 –(23%) swabs, strains related to patient’s contamination – Kocuria kristinae and Corynebacterium – isolated from 3 (8%) swabs, pathogen- ic Gram-negative rods not producing either ESBL or KPC were isolated from 5 (13%) swabs. The remaining strains were classified as opportunistic flora meticilin-resistant S.

hominis, meticilin-resistant S. haemolyticus, meticillin- susceptible S. aureus, E. faecalis, E faecium, E. Hirae and were isolated from 15 (38%) swabs.

The following differences were also observed be- tween centres: bacteria were isolated from 100% of swabs in Centre A, and from 42% in Centre B, of which 80% were taken from equipment in the Radiology De- partment. Bacteria were cultured from 5 out of 8 swabs taken from ultrasound transducers – 3 cultures contained opportunistic strains. In centre B out of 14 swabs taken from scanners located in the Transplantology Depart- ment, Operating Theatre and Intensive Care Unit – ICU bacterial growth was obtained in only 2 samples. Bacte- rial isolates belonged to environmental microflora.

Discussions

There are different types of ultrasound scans: ex- ternal, internal, and endoscopic. Procedures may vary from noninvasive(ex. transabdominal ultrasound, tran- sthoracic echocardiography, small parts and musculo- skeletal examinations, Doppler vascular assessment), through procedures with direct contact with mucous membranes (ex. transesophageal echocardiography, transvaginal ultrasonography), and invasive procedures (ex. endovascular ultrasonography). Sonographers dif- fer in habits and skill during ultrasound scan. The “criti- cal instruments” penetrating the skin or mucous mem- branes require sterilization. Semicritical instruments that come into contact with mucous membranes (e.g.

fiber-optic endoscopes require high-level disinfection).

Noncritical devices come into contact with intact skin but not mucous membranes. External probes that only come into contact with clean, intact skin are considered noncritical devices and require cleaning after every use [5]. These devices should be cleaned by low-level dis- infection techniques [6].

Ultrasound machines and transducers may act as both a source and a vector of nosocomial infection. Several studies have confirmed the transmission of bacteria and viruses from a patient’s skin to ultrasound equipment, the most significant organisms including Staphylococcus au- reus (including methicillin-resistant strains, i.e. MRSA), Pseudomonas, Acinetobacter species, Candida albicans, hepatitis B and C viruses, human immunodeficiency vi-

rus (HIV) and herpes simplex virus, human papilloma virus [7,8].

There are several possible routes of pathogen trans- mission during ultrasound examinations. In order to en- sure good conduction of the ultrasound wave from the transducer to the examined tissues and back again, the skin is covered with a thin layer of gel. The gel is then distributed over the skin of the patient by moving the transducer back and forth. Ultrasound examination is dy- namic: it requires numerous changes in position by both the patient and the examiner. During this process, the ca- ble often comes into contact with the patient skin, and the clothing of the examiner with the gel-covered surface of the skin. After placing the transducer in the holder after use, the possibility of its contamination remains. Most of the radiologist and radiographers are right handed and hold the transducer in right hand. The free hand – usu- ally the left, is used to take measurements, introduce text comments on the obtained images, enable/disable Dop- pler imaging, to adjust the imaging parameters – requir- ing from a few to several movements – by changing the control knob, pressing buttons, selecting fields on the touch screen and using the keyboard controls. In the case of the equipment housing and its controls being dirtied, pathogens can be rapidly transferred to virtually all us- able parts of the equipment.

In many centres, the procedures for preparing the ul- trasound equipment for operation and then leaving it for the following user may become less restrictive moving from the operating theatre and towards the Radiology Department. It is recommended to clean the transducers and the cable after completing each examination with alcohol-free disinfectant. Ethyl and propyl alcohol are effective decontaminants with success rates as high as 100 % [9] however, may damage the transducer, there- fore Koibuchi et al [10] has suggested wiping the probes with a plain paper towel after every ultrasound examina- tion, although to a degree that is not comparable with that obtained by using an alcohol-soaked paper towel.

The other suggestion is wiping the probe with an alcohol- soaked paper towel before examination for patients with immunosuppression disease or after the examination of a patient with a skin infection, and at the end of each work- ing day, additionally, the probe should be wiped with an alcohol-soaked paper towel [11].

According to AIUM (the American Institute of Ul- trasound in Medicine) guidelines transducers should be cleaned after each examination with soap and water or quaternary ammonium (a low-level disinfectant) sprays or wipes. The probes must be disconnected from the ul- trasound scanner for anything more than wiping or spray cleaning. After removal of the probe cover (when applica-

(5)

from the ultrasound probes constitute normal bacterial flora of the skin, and do not pose any threat in case of possible transmission. It is important to note the pres- ence of Gram-negative bacilli on swabs taken from the touch screens of two ultrasound units. The presence of Gram-negative bacteria may be associated with exces- sive humidity in the room where the ultrasound equip- ment is used, faulty ventilation or failure to comply with the appropriate decontamination procedures. In most cases, the touch panel of the equipment is operated with the left hand without the possibility to transfer pathogens from the skin of the person performing examination to transducer. When transmission of pathogens occurs the results of the cultures of swabs taken from touch panel and transducer should be similar.

Contamination by alert pathogens is unacceptable and could be due to the result of incomplete disinfection procedures or the lack of adherence to them. No alert pathogens were isolated during the course of the present study.

According to the recommendations of one of the pro- ducers of the tested ultrasound devices [14], safe and cor- rect operation of the system requires not only the trans- ducer to be cleaned regularly, but the following parts also need to be subjected to weekly cleaning and mainte- nance: LCD display, equipment housing, touch panel and control panel.

In view of the results obtained in the present study, bacterial growth of swabs taken from the touch screen and control panel, suggests the disinfection protocol recommended by the manufacturer for the equipment housing, touch screen and display monitor appears to be insufficient to maintain the desired level of decontami- nation. Information regarding cleaning substances is in- consistent and recommends the use of cleaning agents, disinfectant substances and “domestic disinfectants”.

Analysing the bacterial growth of swabs obtained in centre B there is significant difference in the frequency of positive cultures. Eight out of 10 swabs taken in Ra- diology Department were positive in contrast to swabs taken from scanners located in the Transplantology De- partment, Operating Theatre and Intensive Care Unit (ICU) where bacterial growth was obtained in 2 out of 14. Bacteria belonging to environmental microflora were isolated. In centre A all swabs were positive for bacte- rial growth. Cleaning procedures in both centres were revised and a new checklist for ultrasound room prepa- ration was implemented. These include replacing weak detergent containing wipes with antibacterial disposable wipes for cleaning the surfaces of the ultrasound equip- ment in constant contact with examiner hands as well as gel bottle.

ble), running water should be used to remove any residual gel or debris from the probe, then with a moist gauze pad or other soft cloth and a small amount of mild nonabrasive liquid soap (household dish-washing liquid is ideal) thor- oughly cleanse the probe. Usage of a small brush, espe- cially for crevices and areas of angulation, depending on the design of the particular probe should be considered.

Finally the probe should be rinsed thoroughly with run- ning water, and dried with a soft cloth or paper towel [5].

The issues concerning the preparation of ultrasound equipment to work in conditions where pathogen trans- mission might be possible, are the most rigorously defined in the operating theatre settings. Usually, sterile disposa- ble covers are used for the transducer and the cable, and a sterile plastic cover is placed over the control panel of the equipment. During the examination, sterile ultrasound gel or water or alcoholic solutions are poured directly onto the skin of the patient. These steps do not cause damage to the active surface of the transducer, because it is secured by a protective plastic sleeve. If patients or out-patients ad- mitted to the Radiology Department display skin changes which appear to be infected, it is recommended to spread a thin layer of sterile gel on the skin, and then cover with a thin layer of cling film foil, which effectively prevents contamination of the transducer, or the transducer can be placed in a glove filled with a small amount of gel. Swabs taken from the ultrasound equipment used in the operat- ing theatre in the present study did not demonstrate any bacterial contamination.

Sykes et al [12] report a study where swabs were tak- en from 5 pieces of ultrasound equipment over the course of three months, seeding samples from the keyboard, gel holder, keyboard and gel itself. On 64.5% of samples environmental organisms were cultured, 7.7% potential pathogens, and no bacterial growth was obtained from remaining 27.8 % of samples. Just as in the present study, greater contamination was found to occur in pieces of equipment used for routine examinations as opposed to scanners used in sites requiring higher standards of disin- fection, such as operating theatres.

Taking into account the need for decontamination procedures after each ultrasound examination, the re- sults of Ejtehadi et al [13] are particularly interesting.

The author notes that merely removing the gel from the transducer using a non sterile paper towel is an effective way to decontaminate equipment, and this also applies to transducers coming into contact with infected skin sur- faces and open wounds. Only after the end of the work- ing day were the transducers subjected to decontamina- tion with a chlorhexidine solution.

In our study ultrasound equipment was used for ex- ternal scans only. The bacterial isolates from swabs taken

(6)

In the literature there are papers on a variety of bacte- rial microflora isolated from ultrasound equipment. Our study was conducted as a part of audit activity assuming that personnel adheres to local guidelines on ultrasound equipment decontamination. Diversity of microorgan- isms is related to the epidemiological situation, as well as the carrier status of the person performing the scan and quite probably, that of the patient. Our isolates gen- erally belong to the normal human microflora and strains typically occurring in the environment. Our results con- firm those described by authors from other centres, in- cluding university hospitals [9,12,13,13]. However, in our opinion, special attention should be given to the fact that spore-producing microorganisms were present on a number of surfaces of the ultrasound scanner. There were some limitations of our study – only aerobic cultures were performed thus we were unable to assess the pres- ence of clostridia, for example Clostridium difficile.

In the literature attention is paid to the critical and semicritical elements (for example: the probe) during the performance of ultrasound examination. We have focused on the problem of the possible transmission of strains as- sociated with non-critical sites namely high frequency touch surfaces of ultrasound devices. Transmission of microorganisms, including multidrug resistant strains, that could lead to colonization of the patient, according to our preliminary observation is related to the differ- ences in the scanning technique by individual doctors.

Sonographers differ in habits and skill during ultrasound scan. Furthermore, the presence of an assistant during ex- amination may influence the number of surfaces touched.

This issue is not likely taken into account in the standard procedures for the decontamination of equipment recom- mended by the manufacturers and therefore, the persons responsible for the performance of the decontamination procedure such as technicians, nurses, or in some centers, even doctors (in Poland) will not be aware of these . This issue needs further studies.

Conclusions

Contamination not only affects parts of diagnostic equipment placed in direct contact with the patient, but also, as shown in the study, those surfaces that only medi- cal personnel have had contact with. The isolated strains belong to the human normal and environmental micro- flora as well as opportunistic strains. No alert pathogens were isolated in our study. It is strongly recommended to implement the proper decontamination procedure for ultrasound scanners including, cleaning not only trans- ducers but also the equipment housing the touch panels and control panels with antimicrobial wipes to reduce

possible risk of infection transmission. This study may lead to the limitation of infections resulting from the use of ultrasound scanners.

Conflict of interest: none

References

1. Mettler FA, Bhargavan M, Faulkner K, et al. Radiologic and nuclear medicine studies in the United States and worldwide: frequency, radiation dose, and comparison with other radiation sources. Radiology 2009; 253: 520-531.

2. Westerway S, Basseal J, Brockway A, Hyett J, Carter D.

P25.05: Potential risks associated with an ultrasound ex- amination: a microbial perspective. Ultrasound Obstet Gy- necol 2014; 44 Suppl S1: 332. doi: 10.1002/uog.14483.

3. Gray J, Staerk A, Berchtold M, Hecker W, Neuhaus G, Wirth A. Growth-promoting Properties of Different Solid Nutrient Media Evaluated with Stressed and Unstressed Micro-organisms: Prestudy for the Validation of a Rapid Sterility Test. PDA J Pharm Sci Technol 2010; 64: 249-263.

4. Cowan S, Barrow G, Steel K, Feltham R. Cowan and Steel’s Manual for the Identification of Medical Bacteria.

Cambridge University Press, 2009.

5. American Institute of Ultrasound in Medicine. Guidelines for Cleaning and Preparing External- and Internal-Use Ultrasound Probes Between Patients. Approved 4/2/2014.

http://www.aium.org/officialStatements/57.

6. Shokoohi H, Armstrong P, Tansek R. Emergency depart- ment ultrasound probe infection control: challenges and solutions. Open Access Emerg Med 2015; 7: 1–9.

7. Sahu B, Raine-Fenning N. Ultrasound and the risk of no- socomial cross infection. Ultrasound Obstet Gynecol 2010;

36: 131–133.

8. Oleszkowicz SC, Chittick P, Russo V, Keller P, Sims M, Band J. Infections associated with use of ultrasound trans- mission gel proposed guidelines to minimize risk. Infect Control Hosp Epidemiol 2012; 33: 1235-1237.

9. Tarzani MK, Eshragi N, Asghari BP, Estakhri R, Eshragi A.

Role of ultrasound transducers in transmission of hospital infections. Casp J Intern Med 2010; 1: 134-136.

10. Koibuchi H, Kotani K, Taniguchi N. Ultrasound probes as a possible vector of bacterial transmission. Med Ultrason 2013; 15: 41-44.

11. M’Zali F, Bounizra C, Leroy S, Mekki Y, Quentin-Noury C, Kann M. Persistence of microbial contamination on trans- vaginal ultrasound probes despite low-level disinfection procedure. PLoS One 2014; 9: e93368.

12. Sykes A, Appleby M, Perry J, Gould K. An investigation of the microbiological contamination of ultrasound equip- ment. Br J Infect Control 1996; 7: 16-20.

13. Ejtehadi F, Ejtehadi F, Teb JC, Arasteh MM. A safe and practical decontamination method to reduce the risk of bac- terial colonization of ultrasound transducers. J Clin Ultra- sound 2014; 42: 395-398.

14. Logiq E9 – manual Polish version, GE 3-60,63

Referințe

DOCUMENTE SIMILARE

According to our previous investigations it seems that tolerance, whether regarded as a political practice or a philosophical or moral principle, is a strategy (or tactics) of one

Note: In the sequel, we will consider that the actions are taken in a deterministic way, and show how the prob- lem can be solved. Then we will generalize to the non-

As a result, the new US techniques that can be used for the assessment of prostate disease (contrast-enhanced ultrasound and elastography) and the new role of ultra- sound

There are n + 1 prefixes (including the empty prefix and the com- plete set of points) and each one can either be classified as +1 or as − 1, for a total of at most 2(n +

In terms of sensitivity two studies from the same re- search group comparing US to Magnetic Resonance Imag- ing (MRI) demonstrated that, at shoulder level, US-detect- ed SAD