• Nu S-Au Găsit Rezultate

EchoScopy in scanning abdominal diseases; a prospective single center study

N/A
N/A
Protected

Academic year: 2022

Share "EchoScopy in scanning abdominal diseases; a prospective single center study"

Copied!
8
0
0

Text complet

(1)

DOI:

Original papers

EchoScopy in scanning abdominal diseases; a prospective single center study

Ana Paula Barreiros

1

, Yi Dong

2

, André Ignee

2

, Daniel Wastl

3

, Christoph F. Dietrich

2

1

Deutsche Stiftung Organtransplantation, Region Mitte, Mainz,

2

Medical Clinic 2, Caritas-Krankenhaus Bad Mergen- theim, Academic Teaching Hospital of the University of Würzburg,

3

Medizinische Klinik, Medizinische Intensivstation, Krankenhaus Nordwest Frankfurt am Main, Germany

Received 10.01.2019 Accepted 10.02.2019 Med Ultrason

2019, Vol. 21, No 1, 8-15

Corresponding author: Prof. Dr. med. Christoph F. Dietrich Medizinische Klinik 2, Caritas-Krankenhaus Uhlandstr. 7, 97980 Bad Mergentheim Tel:+49 7931 58 2201

Email: [email protected]

Introduction

The basic ultrasound (US) approach with handheld devices has been defined by the European Federation of Societies of Ultrasound in Medicine and Biology (EFSUMB) as “EchoScopy”, to distinguish the use of EchoScopy from conventional ultrasound and point of care examinations, the latter being US examinations per-

formed bedside with conventional equipment and inter- preted directly by the clinician [1-9]. Early studies were reported on mobile ultrasound equipment in a laptop for- mat with promising results [10-12].

More recently a new category of handheld devices of pocket size similar to a smartphone were developed, therefore, the data are not comparable to the much larger equipment used in earlier studies. Currently the small- est device is the Vscan™ (GE Healthcare) more recently as a dual probe. The EchoScope provides conventional B-mode and colour Doppler imaging (CDI). The com- pact size of the EchoScope makes it possible to carry the ultrasound device almost like a stethoscope under most clinical circumstances including visiting patients [5,13-37]. Owing to this convenience, the EchoScope can be used as an adjunctive tool for physical exami-

Abstract

Background and aims: The introduction of a new type of small handheld ultrasound device brings greater portability and affordability in a different setting. The basic ultrasound approach with these handheld devices has been defined by European Federation of Societies of Ultrasound in Medicine and Biology (EFSUMB) as “EchoScopy”. The current study aimed to as- sess the image quality, indications and limitations of a portable pocket “EchoScopy” performed first compared with a high-end ultrasound system (second) in abdominal diseases.

Material and methods: Three hundred consecutive patients (158 males and 142 females, age 55±19 [18-96]) years) were included. The ultrasound examinations were performed firstly by an EchoScope (Vscan™ Dual Probe) and secondly with a high-end ultrasound system (HEUS, GE Logiq E9). Compared with the always excellent image quality using HEUS, the image quality of the EchoScope was graded as good, sufficient or non-sufficient. Results: Out of all 300 patients, 221 had focal lesions, 31 patients were found with diffuse pathological findings, 20 with ascites, 25 after liver puncture and 45 without any pathological findings. The image quality of the pocket device was considered as being good or sufficient to delineate the pathology in 265/300 (88%). The detection rate of the EchoScope for abdominal focal lesion was 172/221 (78%). The higher frequency of the Dual Probe was helpful in 35/300 (12%). Conclusions: EchoScopy has proven to display sufficient image quality to answer specific questions, e.g., detection of ascites, splenomegaly, bile duct enlargement, hydronephrosis and other pathological findings which can be judged by “yes/no”.

Keywords: guidelines; mobile ultrasound; point of care; EchoScopy

DOI: 10.11152/mu-1907

(2)

nation [8,38]. The objectives of the present prospective study were as follows: To assess the B-mode (including measurements) and colour Doppler image quality of the EchoScope compared with a high end ultrasound system (HEUS), to evaluate the diagnostic accuracy of EchoS- copy in scanning defined abdominal diseases, to evaluate the detection rate of focal liver lesions (FLL) of EchoS- copy, to define and confirm indications for EchoScopy and to identify indications where EchoScopy may not be recommended.

Material and methods Patient recruitment

Three hundred consecutive patients (158 males and 142 females, age 55±19, range 18-96) years were re- cruited requiring an US examination of the abdomen for different reasons. The feasibility study was approved by the institutional board. Informed consent was received from all patients. The investigated organs included the hepatobiliary-pancreatic, gastrointestinal, urogenital and large vessels (aorta) as previously described [2].

Equipment

Two different US systems were used: A Vscan™ Dual Probe pocket device (GE Medical Systems, Milwaukee, WI, USA) and the GE’s Logiq E9 ultrasound system (GE Medical Systems, Milwaukee, WI, USA), an exemplary high-end ultrasound system.

As already mentioned, Vscan™ Dual Probe is a handheld pocket-sized US equipment with unit size of 135x73x28 mm and a total weight of 436 g including a fixed broad-bandwidth phased array transducer (1.7-3.8 MHz) and a second broad-bandwidth linear array trans- ducer (3.4-8.0 MHz) (fig 1). For more comparing figures we refer to the EFSUMB atlas (www.efsumb.org). The size of the display is 3 inches with resolution 240x320 pixels. It has a rechargeable battery with a mean run time

of B-mode images of 90 minutes and B-mode (80%) and CDI (20%) of 60 minutes.

The entire unit including transducer can fit into a white coat pocket and it is designed to be operated with one hand. All the recorded images and MP4 videos are stored in a SD memory card and can be reviewed on its display unit or copied easily to a PC via the included docking station or directly from the SD-card. This device provides conventional B-mode and CDI. Further func- tions such as pulsed wave and continuous wave Doppler are lacking.

The fully equipped high end ultrasound system Logiq E9 allows the visualization of structures regardless of the depth due to its increased penetration and sensitivity.

This is especially important for scanning obese patients and to examine superficially located structures.

Data acquisition and analysis

One experienced physician (Deutsche Gesellschaft für Ultraschall in der Medizin [DEGUM] 3 level, CFD) performed the examinations in this study. With a compre- hensive knowledge of the medical history and symptoms of the patient, the physician performed an US investiga- tion first with the Vscan™ Dual Probe for scanning the abdominal organs and made a diagnosis. Then the US investigation was performed using HEUS. The exami- nation time with Vscan™ Dual Probe and HEUS took 15-30 minutes; the examination time with Vscan™ Dual Probe was slightly shorter since less details were ob- served. Their main diagnoses were compared to inves- tigate the agreement rate as previously described [2]).

Artefacts were considered [39-41]. The thorax, pleura and lung were not included and results were published elsewhere [4,6,39,42,43].

The following abdominal organs were evaluated and listed on-site findings as previously described [2]:

1) liver: focal liver lesion detection [44-47] including focal fatty sparing [48,49]; portal vein (flow direction)

Fig 1. Vscan™ Dual Probe use in scanning the abdomen. Acute biliary pancreatitis is shown (a-c). The gallbladder with stones (a), the bile duct (enlarged) (b), the stone in the common bile duct is assumed (c). A 10 mm lymph node is clearly shown (LN) next to a transmural inflammation in a patient with Crohn’s disease (d).

(3)

[50-53]; detection of complications after biopsy and treatment procedures [54-58]; diffuse liver disease (liver cirrhosis and fatty liver, defined by the detection of focal fatty sparing) [49,59]; 2) biliary system: detection of gall bladder stones, sludge or neoplasia; measurement of bile duct diameter (normal: <6 mm, after cholecystectomy

<10 mm) [60]; cholecystitis; 3) pancreas: detection of fo- cal lesions; diffuse parenchymal disorders with pancreat- ic duct pathology; 3) spleen: splenomegaly (size); lesion detection; lesion characterization; 4) kidney: detection of hydronephrosis; lesion detection including typical cyst;

nephrolithiasis >5 mm; 5) aorta and inferior vena cava:

abdominal aortic aneurysm [15,18,23,37,61-64]; inferior vena cava evaluation [28,36,64]; 6) peritoneal cavity: de- tection of ascites [4,17,23,35]; guiding abdominal para- centesis [57,58,65,66].

For the patients with focal lesions, the largest di- ameters were measured and the colour Doppler signals (blood flow) were graded within the lesion on the same plane with both the Vscan™ Dual Probe and HEUS as previously described [2].

The vascularity in the lesion was classified into 0-III grades [2,51,53]: grade 0: no vessels; grade I: small ves- sels, one or two punctiform or short rod-shaped colour flow signals; grade II: medium sized vessels, one main vessel or ≥3 small vessels; grade III: high vascularity, ≥2 medium sized vessels.

Because the size of the window on CDI cannot be changed on Vscan™, larger lesions may not be totally covered by one colour Doppler window. In this case the window was moved from one side to the other for grading the colour flow comprehensively as previously described [2)]. For some diffuse diseases of abdominal organs, such as fatty liver, the degree of severity was classified as mild, moderate and severe [49,59,67] depending on sonomorphological results.

In addition, the image quality was evaluated: the im- age quality of HEUS was classified as excellent. The im- age quality of the EchoScope Vscan™ Dual Probe was analyzed via display and via PC after copying the images from SD card. They were assessed and classified in three classes: 1: good; 2: sufficient (with noises which did not influence the diagnosis); and 3: insufficient.

Statistical analysis

Continuous data were presented as the mean±standard deviation (SD). The measurements achieved from the Vscan™ Dual Probe and HEUS were compared with paired Student’s t test, and also with the Pearson’s corre- lation test. Wilcoxon rank sum test was used to compare the difference between the grade of colour flow from the Vscan™ Dual Probe and HEUS and also the score of im- age quality, Spearman’s correlation test was applied to

analyze their correlations. p-values <0.05 were consid- ered significant. Tests and calculations were carried out using SPSS package, version 19.0 (SPSS Inc., Chicago, IL, USA).

Results

In 221 patients one or more focal lesions were found;

31 patients showed diffuse parenchymal disorders; in 20 patients ascites could be displayed and 25 patients were investigated after performing a liver biopsy. Fi- nally, 45 patients had no pathological ultrasound find- ings in examination performed with both ultrasound systems.

Image quality

The image quality of the pocket device was consid- ered as being good or sufficient in 265/300 (88%) pa- tients in comparison with the image quality of Logiq E9 as our defined standard of excellence.

The detection rate of the EchoScope for abdominal focal lesions was 172/221 (78%) in comparison to the gold standard. The higher frequency transducer of the Dual Probe was tested in all and helpful in 35/300 (12%).

In addition, the image quality was not sufficient in three patients with appendicitis and 15 patients with pe- ridiverticulitis.

Scanning for focal lesion

In 221 patients, 221 focal lesions were assessed by HEUS in different locations: liver, n=101; gallbladder, n=21; pancreas, n=13; spleen, n=7; kidney, n=30; ab- domen, n=9; aorta, n=7; portal vein, n=2; common bile duct, n=1; colon, n=12; lymph nodes, n=12. EchoScopy was unable to detect the known lesions in 12 patients due to insufficient image quality (four deeply located focal liver lesions, one patient with appendicitis, 2 with diver- ticulitis and five patients with Crohns disease). Taken together, the detection rate of EchoScopy for abdominal focal lesion was 172/221 (78%). The mean diameter of the focal lesions measured with HEUS and EchoScopy were 3.7±2.6 cm and 3.6±2.4 cm, respectively (not sta- tistically significant (p=0.11). CDI was performed in all patients with focal lesions identified both ultrasound sys- tems; 193/221 (88%) of lesions had the same CDI grades on both HEUS and EchoScopy.

Scanning for diffuse disease

Diffuse diseases were diagnosed in 19 patients using

HEUS (fatty liver, n=8; liver cirrhosis, n=5; bowel wall

thickening, n=3; hydronephrosis, n=3). The diagnosis

with EchoScopy and HEUS were in agreement in 18/19

(94.7%) patients. Only one patient with mild fatty liver

disease using HEUS was misdiagnosed as a normal find-

ing on EchoScopy.

(4)

Abdominal paracentesis and assessing complications after intervention

EchoScopy was used for performing abdominal paracentesis, especially to determine the best localiza- tion site for puncture in six patients with ascites. HEUS and EchoScopy had a total (100%, 6/6) agreement for the localization. The depth of the fluid at the position- ing point on HEUS and EchoScopy was 4.03±1.51 cm and 4.08±1.52 cm, respectively. For six patients with un- clear liver tumors who underwent liver biopsy, both US devices showed that there was no hemorrhage or other complication.

Discussions

So far small sized handheld devices have been mainly studied in cardiology (echocardiography) [4,6,33,35,68], vascular indications mainly focusing on abdominal aorta aneurysm [15,18,23,37,61-63] and in the setting of Fo- cused Assessment with Sonography for Trauma (FAST) [4,17,23,35,69-71]. Studies on the use of abdominal dis- eases are sparse [2,22,72-74] and except one [2] mainly focus on case reports [72] and acute cholecystitis patients [74].

The present study analyses the value of a new dual probe ultrasound pocket device in screening abdominal organs compared with a HEUS system (feasibility study).

In slight contrast to the recently published study with ex- aminations of HEUS before echoscopy the results in this study were slightly worse for EchoScopy, explained by the different setting of HEUS before EchoScopy versus EchoScopy before HEUS.

The main results can be summarized as follows: 1) the image quality of EchoScopy was good in compari- son to the excellent image quality of HEUS in examining most abdominal diseases; 2) EchoScopy has an impact for detecting abdominal diseases in certain indications;

3) EchoScopy is not intended to rule out abdominal dis- eases in detail; 4) Vscan™ dual Probe measurements are reliable; 5) EchoScopy has an impact in the localization of the puncture site for paracentesis and assessing com- plications of intervention.

EchoScopy using Vscan™ is hitherto the smallest available handheld device which fits into the pocket of a white coat. In spite of its small size, the image qual- ity is sufficient in many clinical settings. To rule out ab- dominal diseases is not possible [2,7,72,73]. The results are in accordance with other applications, e.g., echocar- diography [4,6,16,17,19,23,24,28,30,32,33,35,75-78]. A reason might be that the quality and the frequency of a probe strongly affect the depth and resolution of the im- age. Due to the fact that the penetration and frequency

of the Vscan™ probe is limited compared to the wider range of HEUS, it determines that the Vscan™, has some deficiencies in scanning superficially and very deep lo- cated organs and structures. One major limitation for most available US equipment is to achieve a good image quality in obese patients. Therefore, it is not surprising that Vscan™ is no exception.

Recently, it could be shown that the image quality via EchoScopy display was significantly better than the im- age quality after transferring and displaying the data via PC. The Vscan™ device is equipped with a 4 GB SD card for the storage of the videos and images, and an im- age is usually 14-20 KB of size, which is much smaller than images of other ultrasound systems. The small size of the image restricts the resolution, and especially the image quality copied on a PC was of poor quality. There was a significant difference between evaluation on the Vscan™ device and after transferring to a PC (p=0.026).

Similar results have also been reported in echocardiog- raphy. We did not exam this factor again, since no other results were expected. The Vscan™ full screen echocar- diography images on PC were of a lesser quality than those from the HEUS equipment.; However, no differ- ence was found between the image quality on the display of Vscan™ device and that of the high-end ultrasound equipment [19]. However, this issue is not of high im- portance for most clinical applications, since diagnosis is often made directly when the examination is finished, but may causes problems when reproducibility is required, for example if interhospital transfer is necessary.

The difference in measurements on the same section by the same sonographer between the two US systems was also estimated which is of importance when it comes to normal reference values [52,60,64]. No significant dif- ferences were found [2]. Nevertheless the measurements intravaginally obtained by Vscan™ with a designed in- travaginal gadget for obstetric and gynecologic lesions were 0.3-0.4 cm lower than those obtained with a high resolution ultrasound device [20], a trend we could ob- serve as well. Although the measurement does not affect the detection of lesion, it may influence the diagnosis of the diseases and predominantly the therapeutic proce- dures [2]. The agreement for abdominal aorta aneurysm diagnosis was high using these two ultrasound systems [15].

CDI using EchoScopy has been established and has

worked well with respect to sensitivity in the abdomen

[2]. It was used in echocardiography in the emergency

setting including trauma [4,6,16,19,33,71-79]. More de-

tails have been recently discussed [2]. The main disad-

vantages and limitations are the small screen size (the

CDI window cannot be changed during Vscan™ exam-

(5)

ination to cover the whole area of a larger lesion) and lack of pulse-wave and continuous Doppler to evaluate the blood flow velocities. It has to be pointed out that CDI plays a crucial role in the differential diagnosis of abdominal focal lesions [2,80-84].

The small size of this miniature scanning device al- lows the physicians to perform US bedside, not only for diagnosis, but also for therapeutic options such as per- forming abdominal paracentesis or detecting complica- tions after puncture. It is beneficial that the patients do not need to be transferred to the US unit and subsequent- ly it simplifies the diagnostic and therapeutic procedures.

According to our experience, Vscan™ can also be used for the puncture of abdominal effusion [2].

Some other studies have shown that the inter-observer agreement was high; both inexperienced and experienced sonographers could produce adequate quality images [18-20]. Another important limitation to this study is that there was no better gold standard than performing an US examination with a HEUS. One additional limitation of this study is that Vscan™ and HEUS examinations were performed by only one physician; therefore, we could not demonstrate the inter-observer variability.

It is expected that the current high frequency probes may facilitate sonographers to examine superficial or- gans, e.g. thyroid, breast and cervical vessels. In general handheld, battery-powered US devices may be very use- ful in a variety of clinical settings. Due to the fact that im- age quality, feasibility, handling and storage options are improving rapidly, Point of Care Ultrasound (POCUS) will achieve a significant importance in outpatient and in- patients’ care, but also in emergency settings. Additional acquisition costs are reasonable.

Conclusion

Our prospective study showed a good feasibility for scanning abdominal diseases with a high detection rate for abdominal pathologies but could not exclude diseases with an appropriate guarantee. Vscan™ Dual Probe does not allow detection very superficially and deep located

lesions. We summarized in Table I the limitations for EcoScopy. The EchoScopy device can be used in the as- sessment of particular defined abdominal diseases under certain clinical settings with yes/no responses. It can also be used, for performing abdominal paracentesis and the evaluation of complications after an abdominal puncture.

Future studies should focus on the point of care value in certain settings, e.g., palliative care and geriatric patients.

Taken all results together, the Vscan™ Dual Probe is a good advancement compared to Vscan™ with only one probe. It is very helpful and can be used in many clinical settings. This mini-handheld ultrasound device improves clinical examination in the every-day setting.

References

1. Piscaglia F, Dietrich CF, Nolsoe C, Gilja OH, Gaitini D.

Birth of “Echoscopy”- The EFSUMB Point of View. Ultra- schall Med 2013;34:92.

2. Barreiros AP, Cui XW, Ignee A, De Molo C, Pirri C, Di- etrich CF. EchoScopy in scanning abdominal diseases: ini- tial clinical experience. Z Gastroenterol 2014;52:269-275.

3. Gilja OH, Piscaglia F, Dietrich C. EFSUMB – European Course Book. Echoscopy – A new concept in mobile ultra- sound. 2014.

4. Wastl D, Helwig K, Dietrich CF. Examination concepts and procedures in emergency ultrasonography. Med Klin Inten- sivmed Notfmed 2015;110:231-239.

5. Dietrich CF. Point of Care Ultrasound (POCUS) using echoscopy - EFSUMB. Available at: http://www.efsumb.

org/guidelines/echoscopy-vscan14012014.pdf 2016.

6. Wastl D, Borgmann T, Helwig K, Dietrich CF. Rapid diag- nostic in the emergency unit: bedside sonography. Dtsch Med Wochenschr 2016;141:317-321.

7. Dietrich CF, Goudie A, Chiorean L, et al. Point of Care Ul- trasound: A WFUMB Position Paper. Ultrasound Med Biol 2017;43:49-58.

8. Moore CL, Copel JA. Point-of-care ultrasonography. N Engl J Med 2011;364:749-757.

9. Nielsen MB, Cantisani V, Sidhu PS, et al. The Use of Hand- held Ultrasound Devices - An EFSUMB Position Paper.

Ultraschall Med 2018 Dec 21. doi:10.1055/a-0783-2303.

10. Seitz K, Vasilakis D, Ziegler M. Efficiency of a portable B-scan ultrasound device in comparison to a high-end ma- chine in abdominal ultrasound. Results of a pilot study. Ul- traschall Med 2003;24:96-100.

11. Ziegler CM, Seitz K, Leicht-Biener U, Mauch M. Detec- tion of therapeutically relevant diagnoses made by sonog- raphy of the upper abdomen: portable versus high-end sonographic units -- a prospective study. Ultraschall Med 2004;25:428-432.

12. Judmaier G, Seitz K. How reliable is sonography of the up- per abdomen with portable sonographic units? What does the future hold? Ultraschall Med 2004;25:408-410.

13. Frederiksen CA, Juhl-Olsen P, Larsen UT, Nielsen DG, Eika B, Sloth E. New pocket echocardiography device Table I. Not recommended abdominal indications for Echo-

Scopy

• Examination of superficially located organs e.g. intestinum, bowel wall estimation

• Detection of superficial pathological findings

• Indications in oncology

• Estimation of vascularity, e.g., inflammatory bowel diseases

• Difficult examination conditions, e.g. obesity

• Indications beyond the definition of EchoSopy

(6)

is interchangeable with high-end portable system when performed by experienced examiners. Acta Anaesthesiol Scand 2010;54:1217-1223.

14. Coskun F, Akinci E, Ceyhan MA, Sahin Kavakli H. Our new stethoscope in the emergency department: handheld ultrasound. Ulus Travma Acil Cerrahi Derg 2011;17:488- 15. Dijos M, Pucheux Y, Lafitte M, et al. Fast track echo of 492.

abdominal aortic aneurysm using a real pocket-ultrasound device at bedside. Echocardiography 2012;29:285-290.

16. Kitada R, Fukuda S, Watanabe H, et al. Diagnostic accuracy and cost-effectiveness of a pocket-sized transthoracic echo- cardiographic imaging device. Clin Cardiol 2013;36:603- 17. Ruddox V, Stokke TM, Edvardsen T, et al. The diagnostic 610.

accuracy of pocket-size cardiac ultrasound performed by unselected residents with minimal training. Int J Cardio- vasc Imaging 2013;29:1749-1757.

18. Siso-Almirall A, Gilabert Sole R, Bru Saumell C, et al.

Feasibility of hand-held-ultrasonography in the screening of abdominal aortic aneurysms and abdominal aortic ath- erosclerosis. Med Clin (Barc) 2013;141:417-422.

19. Testuz A, Muller H, Keller PF, et al. Diagnostic accuracy of pocket-size handheld echocardiographs used by cardi- ologists in the acute care setting. Eur Heart J Cardiovasc Imaging 2013;14:38-42.

20. Troyano Luque JM, Ferrer-Roca O, Barco-Marcellan MJ, Sabatel Lopez R, Perez-Medina T, Perez-Lopez FR. Modi- fication of the hand-held Vscan ultrasound and verification of its performance for transvaginal applications. Ultrason- ics 2013;53:17-22.

21. Bornemann P, Bornemann G. Military family physicians’

perceptions of a pocket point-of-care ultrasound device in clinical practice. Mil Med 2014;179:1474-1477.

22. Lee M, Roberts JM, Chen L, et al. Estimation of spleen size with hand-carried ultrasound. J Ultrasound Med 2014;33:1225-1230.

23. Mancuso FJ, Siqueira VN, Moises VA, et al. Focused Car- diac Ultrasound Using a Pocket-Size Device in the Emer- gency Room. Arq Bras Cardiol 2014;103:530-537.

24. Riley A, Sable C, Prasad A, et al. Utility of hand-held echo- cardiography in outpatient pediatric cardiology manage- ment. Pediatr Cardiol 2014;35:1379-1386.

25. Russell PM, Mallin M, Youngquist ST, Cotton J, Aboul- Hosn N, Dawson M. First “glass” education: telementored cardiac ultrasonography using Google Glass- a pilot study.

Acad Emerg Med 2014;21:1297-1299.

26. Arishenkoff S, Eddy C, Roberts JM, et al. Accuracy of Spleen Measurement by Medical Residents Using Hand- Carried Ultrasound. J Ultrasound Med 2015;34:2203- 2207.

27. Bruns RF, Menegatti CM, Martins WP, Araujo Junior E.

Applicability of pocket ultrasound during the first trimester of pregnancy. Med Ultrason 2015;17:284-288.

28. Dalen H, Gundersen GH, Skjetne K, et al. Feasibility and reliability of pocket-size ultrasound examinations of the pleural cavities and vena cava inferior performed by nurses

in an outpatient heart failure clinic. Eur J Cardiovasc Nurs 2015;14:286-293.

29. Daurat A, Choquet O, Bringuier S, Charbit J, Egan M, Cap- devila X. Diagnosis of Postoperative Urinary Retention Us- ing a Simplified Ultrasound Bladder Measurement. Anesth Analg 2015;120:1033-1038.

30. Lu JC, Sable C, Ensing GJ, et al. Simplified rheumatic heart disease screening criteria for handheld echocardiography. J Am Soc Echocardiogr 2015;28:463-469.

31. Mbuyita S, Tillya R, Godfrey R, Kinyonge I, Shaban J, Mbaruku G. Effects of introducing routinely ultrasound scanning during Ante Natal Care (ANC) clinics on number of visits of ANC and facility delivery: a cohort study. Arch Public Health 2015;73:36.

32. Olesen LL, Andersen A, Thaulow S. Hand-held echocardi- ography is useful for diagnosis of left systolic dysfunction in an elderly population. Dan Med J 2015;62:A5100.

33. Gulic TG, Makuc J, Prosen G, Dinevski D. Pocket-size im- aging device as a screening tool for aortic stenosis. Wien Klin Wochenschr 2016;128:348-353.

34. Johnson GG, Zeiler FA, Unger B, Hansen G, Karakitsos D, Gillman LM. Estimating the accuracy of optic nerve sheath diameter measurement using a pocket-sized, hand- held ultrasound on a simulation model. Crit Ultrasound J 2016;8:18.

35. Colclough A, Nihoyannopoulos P. Pocket-sized point-of- care cardiac ultrasound devices : Role in the emergency department. Herz 2017;42:255-261.

36. Khandwalla RM, Birkeland KT, Zimmer R, et al. Useful- ness of Serial Measurements of Inferior Vena Cava Diam- eter by VscanTM to Identify Patients With Heart Failure at High Risk of Hospitalization. Am J Cardiol 2017;119:1631- 1636.

37. Siso-Almirall A, Kostov B, Navarro Gonzalez M, et al.

Abdominal aortic aneurysm screening program using hand-held ultrasound in primary healthcare. PLoS One 2017;12:e0176877.

38. Dietrich CF, Hoffmann B, Abramowicz J, et al. Medical Student Ultrasound Education: A WFUMB Position Paper, Part I. Ultrasound Med Biol 2019;45:271-281.

39. Dietrich CF, Mathis G, Blaivas M, et al. Lung artefacts and their use. Med Ultrason 2016;18:488-499.

40. Jenssen C, Tuma J, Moller K, et al. Ultrasound artifacts and their diagnostic significance in internal medicine and gas- troenterology - part 2: color and spectral Doppler artifacts.

Z Gastroenterol 2016;54:569-578.

41. Tuma J, Jenssen C, Moller K, et al. [Ultrasound artifacts and their diagnostic significance in internal medicine and gastroenterology - Part 1: B-mode artifacts]. Z Gastroen- terol 2016;54:433-450.

42. Dietrich CF, Mathis G, Cui XW, Ignee A, Hocke M, Hirche TO. Ultrasound of the pleurae and lungs. Ultrasound Med Biol 2015;41:351-365.

43. Dietrich CF, Mathis G, Blaivas M, et al. Lung B-line arte- facts and their use. J Thorac Dis 2016;8:1356-1365.

44. Claudon M, Dietrich CF, Choi BI, et al. Guidelines and good clinical practice recommendations for Contrast En-

(7)

hanced Ultrasound (CEUS) in the liver - update 2012: A WFUMB-EFSUMB initiative in cooperation with repre- sentatives of AFSUMB, AIUM, ASUM, FLAUS and ICUS.

Ultrasound Med Biol 2013;39:187-210.

45. Claudon M, Dietrich CF, Choi BI, et al. Guidelines and good clinical practice recommendations for contrast enhanced ultrasound (CEUS) in the liver--update 2012: a WFUMB- EFSUMB initiative in cooperation with representatives of AFSUMB, AIUM, ASUM, FLAUS and ICUS. Ultraschall Med 2013;34:11-29.

46. Dietrich CF, Jenssen C. Focal liver lesion, incidental find- ing. Dtsch Med Wochenschr 2012;137:2099-2116.

47. Dietrich CF, Sharma M, Gibson RN, Schreiber-Dietrich D, Jenssen C. Fortuitously discovered liver lesions. World J Gastroenterol 2013;19:3173-3188.

48. Jenssen C, Pietsch C, Gottschalk U, et al. Abdominal ultra- sonography in patients with diabetes mellitus. Part 1: Liver.

Z Gastroenterol 2015;53:306-319.

49. Hirche TO, Ignee A, Hirche H, Schneider A, Dietrich CF. Evaluation of hepatic steatosis by ultrasound in pa- tients with chronic hepatitis C virus infection. Liver Int 2007;27:748-757.

50. Dietrich CF, Lee JH, Gottschalk R, et al. Hepatic and portal vein flow pattern in correlation with intrahepatic fat deposi- tion and liver histology in patients with chronic hepatitis C.

AJR Am J Roentgenol 1998;171:437-443.

51. Ignee A, Gebel M, Caspary WF, Dietrich CF. Doppler imag- ing of hepatic vessels - review. Z Gastroenterol 2002;40:21- 52. Sienz M, Ignee A, Dietrich CF.[Reference values in ab-32.

dominal ultrasound - liver and liver vessels. Z Gastroen- terol 2010;48:1141-1152.

53. Ignee A, Boerner N, Bruening A, et al. Duplex sonography of the mesenteric vessels--a critical evaluation of inter-ob- server variability. Z Gastroenterol 2016;54:304-311.

54. Lorentzen T, Nolsoe CP, Ewertsen C, et al. EFSUMB Guide- lines on Interventional Ultrasound (INVUS), Part I. General Aspects (long Version). Ultraschall Med 2015;36:E1-E14.

55. Lorentzen T, Nolsoe CP, Ewertsen C, et al. EFSUMB Guidelines on Interventional Ultrasound (INVUS), Part I. General Aspects (Short Version). Ultraschall Med 2015;36:464-472.

56. Sidhu PS, Brabrand K, Cantisani V, et al. EFSUMB Guide- lines on Interventional Ultrasound (INVUS), Part II. Diag- nostic Ultrasound-Guided Interventional Procedures (Long Version). Ultraschall Med 2015;36:E15-E35.

57. Dietrich CF, Lorentzen T, Appelbaum L, et al. EFSUMB Guidelines on Interventional Ultrasound (INVUS), Part III - Abdominal Treatment Procedures (Long Version). Ultra- schall Med 2016;37:E1-E32.

58. Dietrich CF, Lorentzen T, Appelbaum L, et al. EFSUMB Guidelines on Interventional Ultrasound (INVUS), Part III - Abdominal Treatment Procedures (Short Version). Ultra- schall Med 2016;37:27-45.

59. Dietrich CF, Schall H, Kirchner J, et al. Sonographic detec- tion of focal changes in the liver hilus in patients receiving corticosteroid therapy. Z Gastroenterol 1997;35:1051-1057.

60. Sienz M, Ignee A, Dietrich CF. Reference values in ab- dominal ultrasound - biliopancreatic system and spleen. Z Gastroenterol 2011;49:845-870.

61. Vidakovic R, Schouten O, Feringa HH, et al. Abdominal aortic aneurysm screening using non-imaging hand-held ultrasound volume scanner--a pilot study. Eur J Vasc Endo- vasc Surg 2006;32:615-619.

62. Flu WJ, van Kuijk JP, Merks EJ, et al. Screening for abdom- inal aortic aneurysms using a dedicated portable ultrasound system: early results. Eur J Echocardiogr 2009;10:602-606.

63. Abbas A, Smith A, Cecelja M, Waltham M. Assessment of the accuracy of AortaScan for detection of abdominal aortic aneurysm (AAA). Eur J Vasc Endovasc Surg 2012;43:167- 64. Sienz M, Ignee A, Dietrich CF. Sonography today: refer-170.

ence values in abdominal ultrasound: aorta, inferior vena cava, kidneys. Z Gastroenterol 2012;50:293-315.

65. Sidhu PS, Brabrand K, Cantisani V, et al. EFSUMB Guide- lines on Interventional Ultrasound (INVUS), Part II. Diag- nostic Ultrasound-Guided Interventional Procedures (Long Version). Ultraschall Med 2015;36:E15-E35.

66. Sidhu PS, Brabrand K, Cantisani V, et al. EFSUMB Guide- lines on Interventional Ultrasound (INVUS), Part II. Diag- nostic Ultrasound-Guided Interventional Procedures (Short Version). Ultraschall Med 2015;36:566-580.

67. Dietrich CF, Wehrmann T, Zeuzem S, Braden B, Caspary WF, Lembcke B. Analysis of hepatic echo patterns in chronic hepatitis C. Ultraschall Med 1999;20:9-14.

68. Kimura BJ. Point-of-care cardiac ultrasound techniques in the physical examination: better at the bedside. Heart 2017;103:987-994.

69. Scalea TM, Rodriguez A, Chiu WC, et al. Focused Assess- ment with Sonography for Trauma (FAST): results from an in- ternational consensus conference. J Trauma 1999;46:466-472.

70. Blackbourne LH, Soffer D, McKenney M, et al. Second- ary ultrasound examination increases the sensitivity of the FAST exam in blunt trauma. J Trauma 2004;57:934-938.

71. Kirkpatrick AW, Sirois M, Laupland KB, et al. Hand-held thoracic sonography for detecting post-traumatic pneumo- thoraces: the Extended Focused Assessment with Sonogra- phy for Trauma (EFAST). J Trauma 2004;57:288-295.

72. Alfonzo M, von Reinhart A, Riera A. Point-of-Care Ul- trasound Identification of an Abdominal Hernia. Pediatr Emerg Care 2017;33:596-598.

73. Matthew Fields J, Davis J, Alsup C, et al. Accuracy of Point-of-care Ultrasonography for Diagnosing Acute Ap- pendicitis: A Systematic Review and Meta-analysis. Acad Emerg Med 2017;24:1124-1136.

74. Zenobii MF, Accogli E, Domanico A, Arienti V. Update on bedside ultrasound (US) diagnosis of acute cholecystitis (AC). Intern Emerg Med 2016;11:261-264.

75. Haji DL, Royse A, Royse CF. Review article: Clinical im- pact of non-cardiologist-performed transthoracic echocar- diography in emergency medicine, intensive care medicine and anaesthesia. Emerg Med Australas 2013;25:4-12.

76. Mjolstad OC, Andersen GN, Dalen H, et al. Feasibility and reliability of point-of-care pocket-size echocardiography

(8)

performed by medical residents. Eur Heart J Cardiovasc Imaging 2013;14:1195-1202.

77. Panoulas VF, Daigeler AL, Malaweera AS, et al. Pocket- size hand-held cardiac ultrasound as an adjunct to clinical examination in the hands of medical students and junior doctors. Eur Heart J Cardiovasc Imaging 2013;14:323- 78. Via G, Hussain A, Wells M, et al. International evidence-330.

based recommendations for focused cardiac ultrasound. J Am Soc Echocardiogr 2014;27:683.e1–683.e33.

79. Lin PH, Bush RL, McCoy SA, et al. A prospective study of a hand-held ultrasound device in abdominal aortic aneu- rysm evaluation. Am J Surg 2003;186:455-459.

80. Dietrich CF, Tannapfel A, Jang HJ, Kim TK, Burns PN, Dong Y. Ultrasound Imaging of Hepatocellular Adenoma

Using the New Histology Classification. Ultrasound Med Biol 2019;45:1-10.

81. Dirks K, Calabrese E, Dietrich CF, et al. EFSUMB Position Paper: Recommendations for Gastrointestinal Ultrasound (GIUS) in Acute Appendicitis and Diverticulitis. Ultra- schall Med 2019 Jan 7. doi:10.1055/a-0824-6952.

82. Nurnberg D, Jenssen C, Frohlich E, Ignee A, Mathis G, Di- etrich CF. Ultrasound in palliative care medicine, Part II. Z Gastroenterol 2017;55:582-591.

83. Nurnberg D, Jenssen C, Cui X, Ignee A, Dietrich CF.

Ultrasound in palliative care medicine. Z Gastroenterol 2015;53:409-416.

84. Froehlich E, Pauluschke-Froehlich J, Debove I, Vey W, Klinger C, Dietrich CF. Geriatric ultrasound. Z Gastroen- terol 2017;55:277-290.

Referințe

DOCUMENTE SIMILARE

The Constitution of the Republic of Albania regulates three situations that require extraordinary measures: war situation, state of emergency and state of natural

While textual rhetoric approaches utterances from the viewpoint of their formal organization, interpersonal rhetoric deals with utterances as products of rhetorical

The contribution of this study aims to fuse B-mode and color Doppler modes for breast cancer diagnosis in- cluding (a) feature extraction from two breast US modes (B-mode and

The tribological behavior of different carbonaceous nanomaterials such as single wall carbon nanotubes (SWNTs), functionalized single wall carbon nanotubes (SWNTf), multiwall

Identity is thus constructed in interaction, which means that out of a whole host of potential identity features, those features become salient which permit a differentiation of

The evolution to globalization has been facilitated and amplified by a series of factors: capitals movements arising from the need of covering the external

Using a case study designed for forecasting the educational process in the Petroleum-Gas University, the paper presents the steps that must be followed to realise a Delphi

De¸si ˆın ambele cazuri de mai sus (S ¸si S ′ ) algoritmul Perceptron g˘ ase¸ste un separator liniar pentru datele de intrare, acest fapt nu este garantat ˆın gazul general,