Original papers

DOI: 10.11152/mu-1307

Interconversion of elasticity measurements between two-dimensional shear wave elastography and transient elastography

Yeun-Yoon Kim

1

, Myung-Joon Kim

1

, Hyun Joo Shin

1

, Haesung Yoon

1

, Ha Yan Kim

2

, Mi-Jung Lee

1

1Department of Radiology and Research Institute of Radiological Science, Severance Children’s Hospital, 2Biostatis- tics Collaboration Unit, Yonsei University College of Medicine, Seoul, Korea

Received 31.10.2017 Accepted 28.12.2017 Med Ultrason

2018, Vol. 20, No 2, 127-133

Corresponding author: Mi-Jung Lee, MD, PhD

Department of Radiology and Research Institute of Radiological Science, Severance Children’s Hospital Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemun-gu, Seoul, Korea 03722

Phone: 82-2-2228-7400, Fax: 82-2-393-3035 E-mail: [email protected]

Introduction

Biliary atresia is a major cause of chronic liver dis- ease and an important indication for liver transplantation in children [1-3]. Assessment of hepatic fibrosis is impor- tant for growing children with biliary atresia in terms of disease status evaluation and liver transplantation plan- ning [4-6]. Ultrasound (US) elastography is highly val-

ued for its convenience, non-invasiveness, and reproduc- ibility for the measurement of liver stiffness in adults [7], and therefore can be a powerful tool for long-term fol- low-up evaluation even in pediatric patients. Transient elastography (TE) has been suggested to be useful for assessment of hepatic fibrosis in pediatric patients with chronic liver disease, including biliary atresia [8-11].

Even though there are only a few studies in children, two-dimensional (2D) shear wave elastography (SWE) as well as acoustic radiation force impulse were useful for the differential diagnosis between biliary atresia and other infantile liver diseases [12-14].

Probe selection also should be considered in children.

Various probes are needed for liver stiffness measure- ment (LSM) in children due to their continuous growth and varying body sizes. For TE, the S probe was devel- oped for small children and was previously applied to infants with biliary atresia either before or after Kasai op- Abstract

Aims: To enable comparison and interconversion of elasticity measurements between two-dimensional (2D) shear wave elastography (SWE) and transient elastography (TE). Materials and methods: Elasticities of three phantoms were measured by 2D SWE (supersonic shear imaging) using four probes and TE using two probes. We performed regression analyses to evaluate correlation between the measurements and phantom elasticities, and make converting equations. In pediatric biliary atresia patients who had stiffness measurements by both 2D SWE and TE within 1-year interval, TE measurements were retrospectively converted into correlating 2D SWE values. We compared the calculated values with 2D SWE measurements by intraclass correlation coefficient. Results: Measurements in phantoms varied according to elastography method and probe selection. However, the measurement by both 2D SWE (R2, 0.974-0.985; p<0.001) and TE (R2, 0.996-0.999; p<0.001) showed significant linear correlation with phantom elasticity in all probe settings. From 67 biliary atresia patients (age, 2 months-20 years), agreements between the measured and calculated values were excellent in all 88 examinations within 1-year interval (ρ=0.828; p<0.001) and in 63 examinations within 2-month interval (ρ=0.863, p<0.001). Conclusions: The equations enabled interconversion of elasticity values among different probes of 2D SWE and TE and provided reliable estimation of elasticity values for different probe settings in biliary atresia patients.

Keywords: liver fibrosis; transient elastography; shear wave elastography; biliary atresia; comparative study

(2)

eration [11,15]. For 2D SWE, the small convex probe and high frequency linear probe as well as the large convex probe are available for the evaluation of the liver, which can be especially beneficial to pediatric patients.

The result of LSM in the same individual can be in- fluenced by the probe selection of 2D SWE and TE [16- 18]. Moreover, the shear wave speed measurement can differ according to elastography system made by dif- ferent vendors, including TE and 2D SWE [19]. Con- sequently, appropriate comparison is essential between the results using different systems and probes. In addi- tion, variability of the cut-off values for hepatic fibrosis staging from different systems and probes makes it dif- ficult to directly compare the results of both methods in pediatric patients with chronic liver disease. When using 2D SWE with either SC6-1 or SL15-4 probes for super- sonic shear imaging, for example, cut-off values for any fibrosis (F1 or higher) varied from 7.3 to 10.6 kPa, while 23.5 kPa was suggested as a cut-off value for liver cir- rhosis (F4) [18,20]. For TE using either S or M probes, on the other hand, cut-off value for F1 ranged from 5.1 to 6.1 kPa, while that for F4 ranged from 14.1 to 15.15 kPa [8,11,21]. Therefore, the aim of this study was to en- able comparison and interconversion of elasticity meas- urements between 2D SWE and TE with consideration of variable probe settings.

Materials and methods

We performed a phantom study to establish the rela- tionship between elasticity measurements obtained with 2D SWE and TE which were in the expected range of elasticity values of the patient group. Equations which were made for conversion were retrospectively applied to LSMs of children with biliary atresia for their validation.

Phantom study

We used commercially available phantom models of uniform elasticity (3.0, 16.9, and 26.3 kPa) for elasticity measurements (The Shear Wave Liver Fibrosis Phantom, model 039, CIRS, Norfolk, VA, USA). 2D SWE (Aix- plorer, SuperSonic Imagine, Aix-en-Provence, France) and TE (FibroScan, Echosens, Paris, France) were per- formed in each phantom to measure the elasticity. We put the phantoms in lateral decubitus position and placed probes perpendicularly to the phantom surface in order to simulate intercostal measurement in the real patient.

The following four probes were used for 2D SWE:

SuperCurved probe with 1–6 MHz (SC6-1), SuperLin- ear probe with 2–10 MHz (SL10-2), SuperMicroConvex probe with 3–12 MHz (SC12-3), and SuperLinear probe with 4–15 MHz (SL15-4). Routine pediatric abdominal setting was used for the SC6-1, SL10-2, and SC12-3

probes, and the setting for superficial lesion evaluation was used for the SL15-4 probe. Circular ROIs were placed in the homogeneous areas of the phantoms. The diameter of the ROI was 6 mm except for the SC6-1 probe, for which a ROI diameter of 8 mm was used ac- cording to the routine pediatric abdominal setting. The acquisition depth was 4 cm for the SC6-1 probe, 2 cm for the SL10-2 and SC12-3 probes, and 1.5 cm for the SL15-4 probe, according to the routine setting for each probe. Ten valid measurements from separate and repeti- tive acquisitions were obtained by each probe in the same area of the phantom by a board-certified radiologist. The elasticity measurements were presented as Young modu- lus in the unit of kPa.

For TE, the S2 probe with the diameter of 5 mm and the M probe with the diameter of 7 mm were used. The ultrasonic frequencies were 5 MHz for the S2 probe and 3.5 MHz for the M probe. TE was performed by a trained investigator who was certified by the manufacturer. The median value of ten repetitive and valid measurements was obtained with each probe in each phantom. The TE results were also obtained in the unit of kPa.

Examinations in children with biliary atresia This retrospective analysis on the patient data was performed with the approval of our Institutional Review Board, and the need for informed consent was waived.

All children with biliary atresia who had previously un- derwent a Kasai operation, and had both results of LSM by 2D SWE and TE performed from January 2015 to July 2016 were included as subjects. The time interval be- tween 2D SWE and TE examinations was limited to one year. We reviewed US images and LSM data obtained by 2D SWE and TE. We also reviewed laboratory results regarding liver function and cholestasis. Signs of portal hypertension, such as splenomegaly, ascites or varices, were collected based on the US, clinical or endoscopic results. Body mass index (BMI) at the time of 2D SWE was defined as normal for percentiles lower than 85, overweight for percentiles between 85 and 95, and obese for percentiles of 95 or higher [22,23].

2D SWE was performed with either the SC6-1 or SL10-2 probe through an intercostal space with patients in the supine position and regular free breathing. We used the same routine pediatric abdominal setting as in the phantom study. LSMs were obtained three times in the homogeneous liver parenchyma avoiding vessels and bile ducts, and the mean value of the three elastic- ity measurements was calculated as the representative value [24]. TE was performed with either the S2 or M probe via the intercostal approach. S2 probe was used for children with the thoracic perimeter of 75 cm or less, and M probe was used for children with the thoracic perim-

(3)

Fig 1. A scatter plot of the elasticity values measured by two- dimensional (2D) shear wave elastography (SWE) using four probes (SC6-1, SL10-2, SC12-3, and SL15-4) and transient elastography (TE) using two probes (S2 and M) in elasticity phantom models of 3.0, 16.9, and 26.3 kPa. SC6-1 = Super- Curved probe with 1-6 MHz, SL10-2 = SuperLinear probe with 2-10 MHz, SC12-3 = SuperMicroConvex probe with 3-12 MHz, SL15-4 = SuperLinear probe with 4-15 MHz.

Fig 2. Linear correlations between representative elasticity val- ues in TE and the corresponding elasticity values in 2D SWE with each probe based on the equations from the phantom study: a) Correlations between TE with the S2 probe and 2D SWE with each probe; b) Correlations between TE with the M probe and 2D SWE with each probe.

eter of more than 75 cm. The median value of ten valid measurements was considered to be the representative value. Both methods were conducted in children below three years of age after fasting for four hours and for six hours in older children.

Data and statistical analysis

Linear regression analyses were performed to evalu- ate the correlation between the results of 2D SWE or TE and phantom elasticity for each probe setting. Because unexpectedly we could not obtain elasticity values using either S2 or M probe in the 3.0 kPa phantom, ten values were imputed for missing values in each probe setting by a Markov Chain Monte Carlo method. Linear regres- sion analysis was performed for the data with or without imputed values in TE measurements. Using the equations derived from the linear correlation analyses, the corre- sponding elasticity values in 2D SWE with each probe were calculated from the elasticity values in TE with the S2 and M probe, respectively.

For the patient data, we also calculated the 2D SWE values corresponding to the LSMs by TE using the equa- tions. We then analyzed the agreement between the LSM by 2D SWE and the statistically estimated values for 2D SWE using intraclass correlation coefficient for average measures. Analyses were done both for all examinations (with less than 1-year interval) and for those with less than 2-month interval between 2D SWE and TE. We also analyzed the results in each BMI group.

Statistical Analysis Software (SAS) version 9.2 (SAS Institute Inc., Cary, NC, USA) and SPSS Statistics ver- sion 20.0 (IBM Corporation, Armonk, NY, USA) were used for the analysis. A p value less than 0.05 was con- sidered to indicate statistical significance.

Results

Equations from the phantom study

The results of elasticity measurements by each probe of 2D SWE in 3.0, 16.9, and 26.3 kPa phantoms tended to have larger discrepancies in the stiffer phantoms and to be relatively lower than the actual phantom elastici- ties. However, the results of 2D SWE with all four probes showed significant linear correlation with phantom elas- ticity (R2, 0.974-0.985; p<0.001). The following equa- tions demonstrate the relationships between the phantom elasticity and the results of 2D SWE with each probe.

SC6-1: phantom elasticity = 1.388 + (1.314 × the re- sult of 2D SWE) (R2=0.984)

SL10-2: phantom elasticity = 1.720 + (1.410 × the result of 2D SWE) (R2=0.985)

SC12-3: phantom elasticity = 1.083 + (1.546 × the result of 2D SWE) (R2=0.980)

SL15-4: phantom elasticity = 0.943 + (1.549 × the result of 2D SWE) (R2=0.974)

In TE, the elasticity values for M probe tended to be lower than those for S2 probe for the same correlating phantom elasticity (fig 1). When excluding the imputed values for the 3.0 kPa phantom setting, elasticity val- ues of both the S2 probe (regression coefficient 0.655, R2=0.996, p<0.001) and M probe (regression coefficient 1.011, R2=0.984, p<0.001) demonstrated significant lin- ear correlation with phantom elasticity. When including the imputed values, the correlation was also significant for elasticity values of both the S2 and M probes (R2, 0.996-0.999; p<0.001) and the relationships are demon- strated in the following equations.

S2: phantom elasticity = 10.138 + (0.655 × the result of TE) (R2 = 0.999)

(4)

M: phantom elasticity = 6.445 + (1.102 × the result of TE) (R2 = 0.996)

The difference between measurements by the S probe and M probe also showed a tendency to be larger in the stiffer phantoms and greater in the 26.3 kPa phantom than the differences among measurements by 2D SWE probes. Different linear correlations are demonstrated in fig 2 by correlating representative elasticity values in TE with the S2 probe (fig 2a) or M probe (fig 2b) to the cor- responding elasticity values in 2D SWE with each probe using the aforementioned equations. Representative val- ues in TE with the S2 probe (3.6, 6.7, 9.8, 12.9, and 16 kPa) were calculated from arbitrary representative values

in TE with the M probe (6, 8, 10, 12, and 14 kPa, respec- tively) based on the equations (Table I).

Application of the equations in children with biliary atresia

During the study period, 67 children (27 boys and 40 girls) underwent 88 abdominal US examinations in- cluding 2D SWE and TE within a 1-year interval. The mean time interval between 2D SWE and TE was 62.4 days, and 63 of 88 examinations were performed within a 2-month interval. The age range of patients was 2 months to 20 years, with a mean of 6.6±4.5 years. The majority (80 of 88, 90.9%) of examinations was performed in pa- tients with normal BMI, except for four overweight and Table I. Equations and representative values of phantom elasticity and elasticity measurements

2D SWE TE

Probes SC6-1 SL10-2 SC12-3 SL15-4 S2 M

Equation for each probe

setting 1.388 + (1.314

× the result of 2D SWE)

1.720 + (1.410

× the result of 2D SWE)

1.083 + (1.546

× the result of 2D SWE)

0.943 + (1.549

× the result of 2D SWE)

10.138 + (0.655

× the result of TE)

6.445 + (1.102 × the result of TE) Representative phantom

elasticity value (kPa) Converted elasticity value on each probe setting (kPa)

5.0 2.7 2.3 2.5 2.6 -7.8 -1.3

7.0 4.3 3.7 3.8 3.9 -4.8 0.5

9.0 5.8 5.2 5.1 5.2 -1.7 2.3

11.0 7.3 6.6 6.4 6.5 1.3 4.1

13.0 8.8 8.0 7.7 7.8 4.4 5.9

15.0 10.4 9.4 9.0 9.1 7.4 7.8

20.0 14.2 13.0 12.2 12.3 15.1 12.3

25.0 18.0 16.5 15.5 15.5 22.7 16.8

30.0 21.8 20.1 18.7 18.8 30.3 21.4

35.0 25.6 23.6 21.9 22.0 38.0 25.9

2D SWE – 2 dimensional shear wave elastography; TE – transient elastography Table II. Characteristics of 67 biliary atresia patients

Characteristics Mean ± standard deviations

M:F 27:40

Age at Kasai operation (days) 61.3 ± 28.1

Age at examination (years) 6.6 ± 4.5

Body weight (kg) 26.0 ± 15.3

Body mass index (kg/m2) 17.2 ± 2.7

Laboratory findings at the time of 2D SWE

Total bilirubin (mg/dL) 1.4 ± 2.1

Direct bilirubin (mg/dL) 0.8 ± 1.7

Albumin (g/dL) 4.0 ± 0.5

AST (IU/L) 76.8 ± 83.7

ALT (IU/L) 63.1 ± 70.0

ALP (IU/L) 309.1 ± 153.6

GGT (IU/L) 121.3 ± 163.4

PT-INR 1.12 ± 0.35

Platelet count (109/L) 165.3 ± 88.2

Presence of portal hypertension sign (n, %) 43 (64.2%)

AST – aspartate aminotransferase; ALT – alanine aminotransferase; ALP – alkaline phosphatase; GGT – gamma glutamyl transpeptidase;

PT-INR – prothrombin time-international normalized ratio

(5)

four obese patients. The laboratory data for liver function and cholestasis were mostly stable at the time of exami- nation (Table II). Forty-three (64.2%) of 67 patients had the sign of portal hypertension based on the US, clinical and endoscopic examinations.

For 2D SWE, the SC6-1 probe and SL10-2 probe were used in 72 and 16 examinations, respectively. The LSM ranges of 2D SWE were 4.6 to 35.7 kPa with a mean of 12.1±6.5 kPa in examinations by SC6-1 probe and 6.1 to 29.5 kPa with a mean of 14.9±6.6 kPa in ex- aminations by SL10-2 probe. For TE, the S2 probe and M probe were used in 77 and 11 examinations, respectively.

The LSM ranges of TE were 3.9 to 75.0 kPa with a mean of 15.8±13.0 kPa in examinations by S2 probe and 6.6 to 24.3 kPa with a mean of 11.5±5.1 kPa in examinations by M probe. With respect to the probe selection in 2D SWE and TE, there were 62 examinations by SC6-1 probe and S2 probe, 15 examinations by SL10-2 probe and S2 probe, 10 examinations by SC6-1 probe and M probe, and one examination by SL10-2 probe and M probe.

2D SWE values estimated by our equations from the LSM with TE had excellent agreements with the LSM by 2D SWE both in all 88 examinations within a 1-year in- terval (ρ=0.828, p<0.001) and in 63 examinations within a 2-month interval (ρ=0.863, p<0.001) between 2D SWE and TE (fig 3a). When considering patients’ BMI, 80 ex- aminations of children with normal BMI showed an ex- cellent agreement between the estimated and measured values of 2D SWE (ρ=0.832, p<0.001). The measure- ments in overweight or obese patients (n=8) also had the tendency of agreement with the estimated values of LSM (fig 3b).

Discussions

To the best of our knowledge, this is the first study suggesting equations for converting LSMs by each probe of 2D SWE and TE to the correlating elasticity values in various probe settings. Our results demonstrate that elasticity values differ according to the elastographic methods and probes. There was a tendency of larger discrepancies among the elasticity values in the stiffer phantoms, especially between the results of S probe and M probe in the 26.3 kPa phantom. In TE, the elasticity values for M probe tended to be lower than those for S2 probe for the same correlating phantom elasticity, which was in accordance with previous results [9,15,25]. In 2D SWE, stiffness values for SC6-1 were higher than those for SL15-4 when correlated with our equations, as with the result of the previous study on the pediatric popu- lation [18]. Another phantom study has also shown that SWV measurements by SC6-1 probe had a tendency to

be higher than those by the SL10-2 probe [26]. We as- sume that technical differences such as probe frequency, and vibration amplitude may contribute to the discrep- ancy in values among the probes.

When the equations were applied to the patient data, reliable estimation and conversion of elasticity values were feasible. The 2D SWE values estimated by our equations from TE measurements correlated well with the 2D SWE measurements. In addition, high coefficients were attained both for patients with short and long inter- vals between 2D SWE and TE studies. As most patients had stable liver function and cholestatic level, there may have been small differences in the disease status between examinations with long intervals. The good correlations were also consistently demonstrated in the normal, over- weight and obese patients, which suggest the applicabil- ity of our equations to patients with varied BMIs.

In the phantom study, TE failed to produce any valid measurement in the 3.0 kPa phantom, while 2D SWE successfully obtained elasticity measurements in all phantoms. The measurement failure of TE in the 3.0 kPa phantom could be attributed to relatively low stiffness of the phantom which might have generated signal error or loss. However, measurement value as low as 1.6 kPa has been successfully obtained by TE on polyvinyl alcohol cryogel phantom in a previous study [27]. Although we could not compare the failure rates between 2D SWE and TE in the patient data due to retrospective design, higher failure rate of TE than 2D SWE has been demonstrated Fig 3. Scatter plots of measured and estimated 2D SWE val- ues in children with biliary atresia: a) scatter plot of 2D SWE measurements and statistically estimated 2D SWE values from the TE measurements shows excellent agreement, not only in 63 examinations (ρ=0.863, p<0.001) with less than a 2-month interval (blue circles) but also in all data (ρ = 0.828; p<0.001) including the other 25 examinations with the interval of two months to one year (green circles); b) scatter plot of 2D SWE measurements and statistically estimated 2D SWE values from TE measurements according to the body mass index (BMI).

Eighty examinations of children with normal BMI (<85 per- centiles, blue circles) show excellent agreement (ρ=0.832, p<0.001). Four examinations of overweight children (BMI: 85- 95 percentiles, green circles) and the other four examinations of obese children (BMI >95 percentiles, orange circles) also show the tendency of agreement.

(6)

both in pediatric and adult populations in previous stud- ies [28-30]. Meanwhile, a study with a high prevalence of obesity showed a higher failure rate of 2D SWE com- pared to TE [31]. Achieving reliable measurements also could be affected by the operator experience and patient body habitus [32].

This study has some limitations. First, we did not cor- relate the elastographic results with the pathologic results.

Liver biopsy was not included in the routine management plan for pediatric patients in our institution and was not performed due to its invasiveness and the retrospective nature of our analysis. Future studies which control pa- tient factors influencing the LSM value, including chol- estasis, and use liver biopsy result as a reference standard are warranted to further validate our results. Secondly, we could not apply our equations to the measurements by SC12-3 and SL15-4 probes in the patient data, as these probes were not used in the included patients. However, our study demonstrated excellent agreements using the equations in various probe combinations of 2D SWE (SC6-1 and SL10-2 probes) and TE (S2 and M probes).

Further studies may expand the usage of our equations in other probe combinations. Thirdly, the time interval of 1 year between 2D SWE and TE was relatively wide.

Therefore, we separately performed analysis for exami- nations of a time interval within 2 months and found that examinations having a 2-month interval as well as a 1-year interval had good correlation results. Lastly, we only used the mean or median values for elasticity meas- urements but did not analyze technical quality param- eters. Not only obtaining and comparing values, but also evaluating value reliability is important in quantitative analyses. Additional research is required for this point.

Conclusions

There were significant linear correlations between elasticity measurements by 2D SWE and TE in the phan- tom study. The proposed equations enabled the intercon- version of elasticity values among different probes of 2D SWE and TE and provided reliable estimation of hepatic elasticity values for different probe settings in biliary atresia patients.

Conflict of interest: none

References

1. Chardot C, Carton M, Spire-Bendelac N, Le Pommelet C, Golmard JL, Auvert B. Prognosis of biliary atresia in the era of liver transplantation: French national study from 1986 to 1996. Hepatology 1999;30:606-611.

2. Haafiz AB. Liver fibrosis in biliary atresia. Expert Rev Gas- troenterol Hepatol 2010;4:335-343.

3. Hartley JL, Davenport M, Kelly DA. Biliary atresia. Lancet 2009;374:1704-1713.

4. Hahn SM, Kim S, Park KI, Han SJ, Koh H. Clinical ben- efit of liver stiffness measurement at 3 months after Kasai hepatoportoenterostomy to predict the liver related events in biliary atresia. PLoS One 2013;8:e80652.

5. Chongsrisawat V, Vejapipat P, Siripon N, Poovorawan Y.

Transient elastography for predicting esophageal/gastric varices in children with biliary atresia. BMC Gastroenterol 2011;11:41.

6. Hanquinet S, Courvoisier DS, Rougemont AL, et al. Acous- tic radiation force impulse sonography in assessing children with biliary atresia for liver transplantation. Pediatr Radiol 2016;46:1011-1016.

7. Jeong WK, Lim HK, Lee HK, Jo JM, Kim Y. Principles and clinical application of ultrasound elastography for diffuse liver disease. Ultrasonography 2014;33:149-160.

8. Fitzpatrick E, Quaglia A, Vimalesvaran S, Basso MS, Dha- wan A. Transient elastography is a useful noninvasive tool for the evaluation of fibrosis in paediatric chronic liver dis- ease. J Pediatr Gastroenterol Nutr 2013;56:72-76.

9. Goldschmidt I, Streckenbach C, Dingemann C, et al. Ap- plication and limitations of transient liver elastography in children. J Pediatr Gastroenterol Nutr 2013;57:109-113.

10. de Ledinghen V, Le Bail B, Rebouissoux L, et al. Liver stiff- ness measurement in children using FibroScan: feasibility study and comparison with Fibrotest, aspartate transami- nase to platelets ratio index, and liver biopsy. J Pediatr Gas- troenterol Nutr 2007;45:443-450.

11. Shen QL, Chen YJ, Wang ZM, et al. Assessment of liver fibrosis by Fibroscan as compared to liver biopsy in biliary atresia. World J Gastroenterol 2015;21:6931-6936.

12. Leschied JR, Dillman JR, Bilhartz J, Heider A, Smith EA, Lopez MJ. Shear wave elastography helps differentiate biliary atresia from other neonatal/infantile liver diseases.

Pediatr Radiol 2015;45:366-375.

13. Wang X, Qian L, Jia L, et al. Utility of Shear Wave Elas- tography for Differentiating Biliary Atresia From Infantile Hepatitis Syndrome. J Ultrasound Med 2016;35:1475- 1479.

14. Hanquinet S, Courvoisier DS, Rougemont AL, et al. Contri- bution of acoustic radiation force impulse (ARFI) elastog- raphy to the ultrasound diagnosis of biliary atresia. Pediatr Radiol 2015;45:1489-1495.

15. Kim S, Kang Y, Lee MJ, Kim MJ, Han SJ, Koh H. Points to be considered when applying FibroScan S probe in children with biliary atresia. J Pediatr Gastroenterol Nutr 2014;59:624-628.

16. Myers RP, Pomier-Layrargues G, Kirsch R, et al. Feasibil- ity and diagnostic performance of the FibroScan XL probe for liver stiffness measurement in overweight and obese pa- tients. Hepatology 2012;55:199-208.

17. Ferraioli G, Lissandrin R, Zicchetti M, Filice C. Assessment of liver stiffness with transient elastography by using S and M probes in healthy children. Eur J Pediatr 2012;171:1415.

(7)

18. Franchi-Abella S, Corno L, Gonzales E, et al. Feasibility and Diagnostic Accuracy of Supersonic Shear-Wave Elas- tography for the Assessment of Liver Stiffness and Liver Fibrosis in Children: A Pilot Study of 96 Patients. Radiol- ogy 2016;278:554-562.

19. Hall TJ, Milkowski A, Garra B, et al. RSNA/QIBA: Shear wave speed as a biomarker for liver fibrosis staging. Ultra- sonics Symposium (IUS) 2013:397-400.

20. Tutar O, Beser OF, Adaletli I, et al. Shear wave elastogra- phy in the evaluation of liver fibrosis in children. J Pediatr Gastroenterol Nutr 2014;58:750-755.

21. Nobili V, Vizzutti F, Arena U, et al. Accuracy and repro- ducibility of transient elastography for the diagnosis of fi- brosis in pediatric nonalcoholic steatohepatitis. Hepatology 2008;48:442-448.

22. Cole TJ, Bellizzi MC, Flegal KM, Dietz WH. Establishing a standard definition for child overweight and obesity world- wide: international survey. BMJ 2000;320:1240-1243.

23. Park HW, Yoo HY, Kim CH, et al. Reference values of body composition indices: the Korean National Health and Nutri- tion Examination Surveys. Yonsei Med J 2015;56:95-102.

24. Shin HJ, Kim MJ, Kim HY, Roh YH, Lee MJ. Optimal Ac- quisition Number for Hepatic Shear Wave Velocity Meas- urements in Children. Plos One 2016;11:e0168758.

25. Pradhan F, Ladak F, Tracey J, Crotty P, Myers RP. Feasi- bility and reliability of the FibroScan S2 (pediatric) probe compared with the M probe for liver stiffness measurement in small adults with chronic liver disease. Ann Hepatol 2013;12:100-107.

26. Shin HJ, Kim MJ, Kim HY, Roh YH, Lee MJ. Compari- son of shear wave velocities on ultrasound elastography between different machines, transducers, and acquisition depths: a phantom study. Eur Radiol 2016;26:3361-3367.

27. Cournane S, Cannon L, Browne JE, Fagan AJ. Assess- ment of the accuracy of an ultrasound elastography liver scanning system using a PVA-cryogel phantom with op- timal acoustic and mechanical properties. Phys Med Biol 2010;55:5965-5983.

28. Leung VY, Shen J, Wong VW, et al. Quantitative elastogra- phy of liver fibrosis and spleen stiffness in chronic hepati- tis B carriers: comparison of shear-wave elastography and transient elastography with liver biopsy correlation. Radiol- ogy 2013;269:910-918.

29. Bota S, Paternostro R, Etschmaier A, et al. Performance of 2-D Shear Wave Elastography in Liver Fibrosis Assessment Compared with Serologic Tests and Transient Elastography in Clinical Routine. Ultrasound Med Biol 2015;41:2340-2349.

30. Engelmann G, Gebhardt C, Wenning D, et al. Feasibil- ity study and control values of transient elastography in healthy children. Eur J Pediatr 2012;171:353-360.

31. Cassinotto C, Lapuyade B, Mouries A, et al. Non-invasive assessment of liver fibrosis with impulse elastography:

comparison of Supersonic Shear Imaging with ARFI and FibroScan(R). J Hepatol 2014;61:550-557.

32. Gradinaru-Tascau O, Sporea I, Bota S, et al. Does expe- rience play a role in the ability to perform liver stiffness measurements by means of supersonic shear imaging (SSI)? Med Ultrason 2013;15:180-183.

Referințe

DOCUMENTE SIMILARE

A recent multicenter study [18] validated S-shear- wave elastography (Samsung RS80A® ultrasound sys- tem Samsung Medison) in the assessment of liver fibro- sis taking TE

Aim: To evaluate the ultrasound (US) modifications [grey scale, Doppler, 2D-share wave elastography (2D-SWE)] of salivary (parotid and submandibular) and lacrimal glands in

Aims: To compare the effects of 2 and 5 min of passive static stretching (SS) on stiffness and blood flow in the rectus femoris in adolescent athletes using shear wave

The results are expressed as percent (exams with technical success and reliable measurements/ all exams) ; †Comparison of applicability rates between four elastography platforms;

The purpose of this study was thus to compare the Intra-/Inter-observer variation and accuracy in vitro of point shear wave elastography (pSWE) with that of 2D-SWE and to

Aims: To evaluate the differences and changes of testicular volume and elasticity in normal (NL) testes and undescended testes (UDTs) of children using shear wave elastography

Aims: The aim of this study was to prospectively assess the feasibility of the Virtual Touch tissue imaging quantification (VTIQ) method of shear wave elastography (SWE) for

Mate- rial and methods: Small, medium and large ROI sizes in addition to longitudinal, oblique and transverse orientations over five different locations within the rectus